background image

 

 

OEM MOPA Light Sources. User Manual.

 

 

Document ID:

 SL.3328.01.000RE / Rev. 1.0, Date: 25/09/2020                                                                                      Page 

13

 of 

23

 

SUPERLUM Diodes Ltd., Ireland                                                                                                               

www.superlumdiodes.com

 

 

DO NOT

 remove the optical dust cap until a clean FC/APC connector is immediately ready to insert into the 

unprotected optical output port. 

 

IMPORTANT!

 

 
 

If discrepancies are found when comparing the observed optical output or spectrum to the ATR please contact 
SUPERLUM  before  further  using  the  Light  Source.  Using  the  lightsource  without  identifying  the  cause  of  the 
observed discrepancies may result in a reduction of the expected lifetime of the device. 

 

The  user  should  regularly  compare  actual  device  performances  to  those  originally  specified  in  ATR 
while using the Light Source. 

 

Содержание OEM-MOPA

Страница 1: ...EM Master Oscillator Power Amplifier SLD Light Sources User Manual ATTENTION ELECTROSTATIC SENSITIVE DEVICES SUPERLUM Diodes Ltd 2016 2020 All rights reserved B3 Fota Point Enterprise Park Carrigtwohi...

Страница 2: ...SS 3R to 3B of IEC 60825 1 For further information regarding laser safety and laser classification of the Light Source please contact SUPERLUM Warranty Superlum products are guaranteed for a period of...

Страница 3: ...ontrol Features 7 2 Technical Specification 8 2 1 Electrical Connections 8 2 2 General Specification 10 2 3 Fiber and Optical Connection Specification 10 2 4 Other Electro Optical Characteristics 10 3...

Страница 4: ...mote port interfaces as standard options Please note that the OEM MOPA light source is intended for integration into the end user s equipment and is to be considered as an optical component For this r...

Страница 5: ...kage Contents The OEM MOPA light source is coming with the following accessories as standard ones 1 x connector for the DC Power input port 1 x connector for the remote input port 1 x flat cable 0 5 m...

Страница 6: ...er Manual Document ID SL 3328 01 000RE Rev 1 0 Date 25 09 2020 Page 6 of 23 SUPERLUM Diodes Ltd Ireland www superlumdiodes com 1 4 Outlines and Dimensions Figure 1 Drawings of Light Source Dimensions...

Страница 7: ...al operation is disabled USB SERVICE mode LED reflects the status of the combined optical output i e all active SLD channels together that user pre selected in the companion software or all channels a...

Страница 8: ...3 below There are three separate connectors ports first for 12V DC power supply second for REMOTE control external switching of SLDs on and off by analog signal triggering and remote interlock control...

Страница 9: ...n to switching optical output ON 8 OUTPUT ON OFF P IN Allows switching the optical output Apply 5VDC voltage from external power supply between this pin and any LOGIC GND pin to switching optical outp...

Страница 10: ...ble be less than 3m When installing the Light Source into an application where the 12 V DC power cable is over 3m Superlum advise that a ferrite of the type W rth 742 701 5 with 6 loops or similar is...

Страница 11: ...e buyer user to design and certify any equipment incorporating any OEM source of Superlum to meet all local safety regulations prior to sale to the public The texts of these regulations are available...

Страница 12: ...ation sensitive OSAs or OSAs with MM fiber optical inputs may disturb the measurement results i e introduce misbalances distortions and artifacts to measured spectral shapes 4 1 5 If no abnormality wa...

Страница 13: ...ected optical output port IMPORTANT If discrepancies are found when comparing the observed optical output or spectrum to the ATR please contact SUPERLUM before further using the Light Source Using the...

Страница 14: ...t or 2 GB 64 bit for Windows Vista and Windows 7 Processor 100 megahertz MHz or faster processor for Windows 9x and ME 300 MHz or faster processor for Windows 2000 and XP 1 gigahertz GHz or faster 32...

Страница 15: ...Source 4 3 4 Brief description of software operation Run the software When starting the software for the first time the following window for selecting of COM port will appear If the drop down list fo...

Страница 16: ...fore saving and e mailing the generated self test file The SETTINGS area of the software display allows the following control features Activate deactivate the remote interlock The interlock is activat...

Страница 17: ...click USB MODE display button to toggle computer control on and off Manual control of the Light Source from the front device panel becomes disabled when the REMOTE CONTROL box has been checked on the...

Страница 18: ...the process of device initialization assigning COM port It may take few minutes until the process will be completed Corresponding icon of the installer must appear in the system tray Upon completion o...

Страница 19: ...ror message CR LF line termination sequence CR ASCII13 LF ASCII10 TYPE Light Source type five ASCII printable chars MOPA two channels Light Source VH Major part of firmware version 0 9 one ASCII chars...

Страница 20: ...w for signal identification SIGNAL BIT DESCRIPTION OME 0 1 Optical module enabled 0 Optical module disabled TON 1 1 TEC ON 0 TEC OFF TGD 2 1 Optical module temperature stabilized 0 Optical module temp...

Страница 21: ...optical module parameter read data from ADC Request U M CH PN CR LF Acknowledge U M CH PN ADATA CR LF Error message E CR LF common error message M CR LF wrong mode set message CH Channel number 1 2 o...

Страница 22: ...0 ohm FFFF 65534 ohm 3 Optical module maximum current value 0000 270F four ASCII chars 0000 0 0 mA 270F 999 9 mA 4 Optical module current value 0000 270F four ASCII chars 0000 0 0 mA 270F 999 9 mA 5...

Страница 23: ...Error message E CR LF common error message M CR LF wrong mode set message MN Optical module number 1 or 2 One ASCII character PN Parameter number 5 One ASCII character DATA Parameter data See table 2...