OEM MOPA Light Sources. User Manual.

 

 

Document ID:

 SL.3328.01.000RE / Rev. 1.0, Date: 25/09/2020                                                                                      Page 

12

 of 

23

 

SUPERLUM Diodes Ltd., Ireland                                                                                                               

www.superlumdiodes.com

 

 

4 Installation and Using the Light Source

 

 

4.1 Getting Started 

 
It is recommended that the following procedures are completed prior to using the lightsource for the first time.  
 

4.1.1

  Fix  the  Light  Source  onto  an  appropriate  heat  sink  firmly  (refer  also  to  Sec.  3.1  of  the  document  for 

more details).  

 

4.1.2

  Apply  12  V  DC  power.  Ensure  that  the  DC  power  supply  unit  in  use  fulfills  the  requirements  of  the 

“Absolute  Maximum  Ratings”  (Table  4)  above,  including  the  ripple  requirement.  Do  not  make  other 
connections at this stage (USB, etc.). 
 
The 

“STATUS” LED on the device panel will flicker for a short time on startup (for approximately 15 sec), and 

will light constant green when startup period is complete. 
 

4.1.3

 Push the 

“ENABLE” button on the front panel to switch the optical output on – the “OUTPUT” LED will 

then light green (a short delay is acceptable due to the soft start of the optical output). All SLDs installed will 
be powered on. 
 
Measure output power with a dedicated power-meter, and compare the result with the power value specified 
in  the  individual  Acceptance  Test  Report  sheet  provided  (ATR).  Make  sure  that  output  power  you  obtain  is 
comparable to that on the datasheet i.e. within 

±10%. 

 

4.1.4

  Connect  the  light  source  to  a  spectrum  analyzer.  Compare  the  spectrum  with  that  depicted  in  ATR. 

There  should  be  no  perceptible  differences  between  the  two  spectrums.  Polarization  insensitive  spectrum 
analyzers  (OSAs),  e.g.  ANDO  AQ6317B  /  Yokogawa  AQ6370  OSAs  are  recommended  for  this  task. 
Polarization sensitive OSAs, or OSAs with MM fiber optical inputs may disturb the measurement results, i.e. 
introduce misbalances, distortions and artifacts to measured spectral shapes. 

 

 

4.1.5

 If no abnormality was found on the preceding stages the lightsource is now ready for use also when the 

Light Source has been run successfully for the first time it can now be controlled via the USB port, or analog 
signals via the remote port (see Fig. 1, 2, and Sec. 4.4). To prolong the device lifetime, it is recommended to 
switch off the optical output when the lightsource is not in use. 
 

WARNING 

Optical output in fibre-coupled light sources is achieved by the use of an FC /APC mating sleeve and FC/APC 
connectors (optical output port). FC/APC connectors are sensitive to even the slightest contamination. Failure 
to  comply  with  the  following  operating  requirements  may  result  in  contamination  entering  the  optical  output 
port severely reducing the performance of the unit and may result in permanent damage to the optical output 
port.  Any such occurrence is the responsibility of the user and is 

NOT covered by the warranty

DO NOT:

 Insert any FC/APC connector into the optical output port without first ensuring it is clean. It is vital 

that  you  clean  the  fiber  ferrule  before  each  insertion.

  No  known  cleaning  methods  are  100%  effective 

therefore it is imperative that inspection is included as part of the cleaning process.

 

Cleaning  FC/APC  connectors:  using  a  lint  free  cloth  and  isopropyl  alcohol  (Min  99%)  and  clean 
compressed air to dry off the connector is recommended.

 

 

Cleaning the optical output port: Using an Improper cleaning method may result in permanent damage to 
the optical output port. ONLY users who have the knowledge, training and equipment should attempt cleaning 
the optical output port. For further information please contact Superlum 

Содержание OEM-MOPA

Страница 1: ...EM Master Oscillator Power Amplifier SLD Light Sources User Manual ATTENTION ELECTROSTATIC SENSITIVE DEVICES SUPERLUM Diodes Ltd 2016 2020 All rights reserved B3 Fota Point Enterprise Park Carrigtwohi...

Страница 2: ...SS 3R to 3B of IEC 60825 1 For further information regarding laser safety and laser classification of the Light Source please contact SUPERLUM Warranty Superlum products are guaranteed for a period of...

Страница 3: ...ontrol Features 7 2 Technical Specification 8 2 1 Electrical Connections 8 2 2 General Specification 10 2 3 Fiber and Optical Connection Specification 10 2 4 Other Electro Optical Characteristics 10 3...

Страница 4: ...mote port interfaces as standard options Please note that the OEM MOPA light source is intended for integration into the end user s equipment and is to be considered as an optical component For this r...

Страница 5: ...kage Contents The OEM MOPA light source is coming with the following accessories as standard ones 1 x connector for the DC Power input port 1 x connector for the remote input port 1 x flat cable 0 5 m...

Страница 6: ...er Manual Document ID SL 3328 01 000RE Rev 1 0 Date 25 09 2020 Page 6 of 23 SUPERLUM Diodes Ltd Ireland www superlumdiodes com 1 4 Outlines and Dimensions Figure 1 Drawings of Light Source Dimensions...

Страница 7: ...al operation is disabled USB SERVICE mode LED reflects the status of the combined optical output i e all active SLD channels together that user pre selected in the companion software or all channels a...

Страница 8: ...3 below There are three separate connectors ports first for 12V DC power supply second for REMOTE control external switching of SLDs on and off by analog signal triggering and remote interlock control...

Страница 9: ...n to switching optical output ON 8 OUTPUT ON OFF P IN Allows switching the optical output Apply 5VDC voltage from external power supply between this pin and any LOGIC GND pin to switching optical outp...

Страница 10: ...ble be less than 3m When installing the Light Source into an application where the 12 V DC power cable is over 3m Superlum advise that a ferrite of the type W rth 742 701 5 with 6 loops or similar is...

Страница 11: ...e buyer user to design and certify any equipment incorporating any OEM source of Superlum to meet all local safety regulations prior to sale to the public The texts of these regulations are available...

Страница 12: ...ation sensitive OSAs or OSAs with MM fiber optical inputs may disturb the measurement results i e introduce misbalances distortions and artifacts to measured spectral shapes 4 1 5 If no abnormality wa...

Страница 13: ...ected optical output port IMPORTANT If discrepancies are found when comparing the observed optical output or spectrum to the ATR please contact SUPERLUM before further using the Light Source Using the...

Страница 14: ...t or 2 GB 64 bit for Windows Vista and Windows 7 Processor 100 megahertz MHz or faster processor for Windows 9x and ME 300 MHz or faster processor for Windows 2000 and XP 1 gigahertz GHz or faster 32...

Страница 15: ...Source 4 3 4 Brief description of software operation Run the software When starting the software for the first time the following window for selecting of COM port will appear If the drop down list fo...

Страница 16: ...fore saving and e mailing the generated self test file The SETTINGS area of the software display allows the following control features Activate deactivate the remote interlock The interlock is activat...

Страница 17: ...click USB MODE display button to toggle computer control on and off Manual control of the Light Source from the front device panel becomes disabled when the REMOTE CONTROL box has been checked on the...

Страница 18: ...the process of device initialization assigning COM port It may take few minutes until the process will be completed Corresponding icon of the installer must appear in the system tray Upon completion o...

Страница 19: ...ror message CR LF line termination sequence CR ASCII13 LF ASCII10 TYPE Light Source type five ASCII printable chars MOPA two channels Light Source VH Major part of firmware version 0 9 one ASCII chars...

Страница 20: ...w for signal identification SIGNAL BIT DESCRIPTION OME 0 1 Optical module enabled 0 Optical module disabled TON 1 1 TEC ON 0 TEC OFF TGD 2 1 Optical module temperature stabilized 0 Optical module temp...

Страница 21: ...optical module parameter read data from ADC Request U M CH PN CR LF Acknowledge U M CH PN ADATA CR LF Error message E CR LF common error message M CR LF wrong mode set message CH Channel number 1 2 o...

Страница 22: ...0 ohm FFFF 65534 ohm 3 Optical module maximum current value 0000 270F four ASCII chars 0000 0 0 mA 270F 999 9 mA 4 Optical module current value 0000 270F four ASCII chars 0000 0 0 mA 270F 999 9 mA 5...

Страница 23: ...Error message E CR LF common error message M CR LF wrong mode set message MN Optical module number 1 or 2 One ASCII character PN Parameter number 5 One ASCII character DATA Parameter data See table 2...

Отзывы: