3-Gas Analyzer PGA3500 

Revision Level - D 

with an instrument that can provide real-time temperature data.  The purpose of this function 

is to prevent the instrument from pulling a bad sample, which could potentially damage the 

sensors.  The minimum temperature should always be slightly higher than the lowest possible 

process temperature.  The range of values for the Minimum Temp is 0 to 2000.   

The Minimum mV is the millivolt set point that will stop the COF/PF adjustment. This will 

prevent adjustments from being made when the proper conditions are not met. The minimum 

millivolts set point should be slightly higher than the minimum millivoltage that is expected.  

The range of values for the Minimum mV is 0 to 2000.   

The Maximum Adjustment is the largest increment of change that will be applied to automatic 

COF/PF adjustments. This will dictate the size of the steps the instrument takes when it 

adjusts the COF or PF of the control instrument. A higher value will result in more speed in 

achieving the desired set point, but a lower value will result in a smoother approach. The 

Maximum Adjustment feature should be coordinated with the Update Interval time to achieve 

the desired responsiveness.  The range of values for the Maximum Adjustment is 0 to 20

The Maximum COF/PF establishes the upper allowable limit for the COF/PF. For example, if 

this is set to 300, then the instrument will make changes to the COF/PF until it is at 300, but 

it will not go any higher. The value will range from 0 to 999

The Minimum COF/PF establishes the lower allowable limit for the COF/PF. For example, if 

this is set to 100, then the instrument will make changes to the COF/PF until it is at 100, but 

it will not go any lower. The value will range from 0 to 999

The IR Mode is a selection to determine if the instrument is running in Automatic or Manual 

mode. When in Monitor mode, no changes will be made to the COF/PF of the control 

instrument. There are two Automatic modes to select from. COF/PF Adj. %C will make 

adjustments to the COF/PF in the atmosphere controller based on the calculated percent 

carbon. This is the preferred method of adjustment. The other Automatic mode, COF/PF 

adj. CO, will make adjustments based only on the measured level of CO and not on the 

carbon calculation.  The three options are: 

Monitor 

COF/PF Adj. %C 

COF/PF adj. CO 

To cycle between each selection, press the circular arrow key or enter the number that 

corresponds to the option’s position in the list and press Enter, where Monitor is at position 0 

and COF/PF adj. CO is at position 2

The On Delay determines the amount of time it takes for the pump to turn on after both of 

the operating parameters (temperature and millivolts) have been met. The value will range 

from 0 to 60 minutes. 

The Off Delay determines the amount of time it takes for the pump to turn off after either of 

the operating parameters (temperature and millivolts) falls outside of the pre-defined 

boundaries. The value will range from 0 to 60 minutes. 

The Update Interval determines the amount of time between automatic updates of the 

COF/PF on the control instrument. This is adjustable in one-minute increments. It is 

recommended that the Update Interval be set to one (1) minute, and the Maximum 

Adjustment set to one (1) point. This will allow the COF/PF to be adjusted frequently, but 

only a small amount at a time, which will result in smooth operation. The value will range 

from 0 to 20

The Sample Delay determines the amount of time the PGA will wait to continue to adjustment 

of the COF or the PF once the pump has turned on.  The range of values is 0 to 30

Example – The following will show how to set up the sampling parameters of:  

Minimum Temp – 1500  

Содержание PGA 3500

Страница 1: ...ns before operating this equipment Super Systems Inc is not responsible for damages incurred due to a failure to comply with these instructions If at any time there are questions regarding the proper...

Страница 2: ...Display Values Menu Page 5 12 Help Page Menu Page 6 13 Calibration Dates and Run Times Menu Page 7 13 Language Lengua Menu Page 8 13 Battery Status Menu Page 9 14 About Sign On Menu Page 10 14 Revisi...

Страница 3: ...ng reshipment Super Systems will not be responsible for damage caused by improper packaging Moisture and Condensation Depending on the composition of the gas being sampled it is possible that condensa...

Страница 4: ...ent Measuring the level of CO in the carrier gas allows correction of the carbon reading at the furnace Heat Treat Furnaces Conventional Endo Gas Furnace atmosphere carbon potential C can be verified...

Страница 5: ...e CO and CH4 sensors have been calibrated to be most accurate within the ranges as shown above however both sensors are capable of measuring gas concentrations of up to 100 00 Sampling method Extracti...

Страница 6: ...uously pressed toggles between the main page and the menu list On most pages this key can be used to turn the pump on and off without using Screen 4 On some screens it is used to change a selection or...

Страница 7: ...ure smooth operation When new the elements in both of these filters are a yellow white color Both are oriented in the sample stream in a manner that causes any contaminants to collect on the outside o...

Страница 8: ...SET DISPLAY VALUES 6 HELP 7 CALIBRATION DATES 8 LANGUAGE LENGUA 9 BATTERY STATUS 10 ABOUT SIGN ON 11 REVISION DISPLAY 12 LOGGED DATA START DATE 13 LOGGED DATA DISPLAY 14 SESSION 15 SAMPLING PARAMETERS...

Страница 9: ...and also displays the internal temperature IT of the instrument This internal temperature should never exceed 50 Note about Menu Numbers Each menu screen has a unique number that will be displayed in...

Страница 10: ...area up and down the following data should be entered FC TC This is the furnace thermocouple value or the furnace temperature PB MV This is the millivoltage from the carbon probe PB TC This is the pro...

Страница 11: ...s not to calculate the probe carbon but to enable the PGA to compute the suggested COF PF Without knowing the current instrument values it cannot compute the suggested values What is a CO Factor or a...

Страница 12: ...set the pump mode The sample pump mode can be changed from OFF to ON to AUTO by pressing the circular arrow key Auto mode will turn the pump on when the flow is below approximately 1 5 SCFH and off wh...

Страница 13: ...hlight the desired location by using the up and down arrow keys Then enter the number using the numeric keypad and pressing Enter Pressing the ESC button once will clear any typed data that is on the...

Страница 14: ...recharge the battery but it will take a few days to accomplish this By leaving the instrument plugged in and periodically checking the battery voltage to see if it is increasing you will be able to de...

Страница 15: ...date Logged Data File Menu Page 13 Each minute there are 15 points of data collected The time of each data point is shown along the left hand side of the screen The arrow buttons will scroll backwards...

Страница 16: ...art a session on the PGA A session will allow the user to identify the operation of the PGA and break all of the logged data into manageable sections Sessions can be viewed on a computer with the XGA...

Страница 17: ...ocations can all be modified and added through the XGA Viewer software Once a session has started the second page of the session menu will be displayed From this page the user can add a note to the se...

Страница 18: ...range from 0 to 999 The IR Mode is a selection to determine if the instrument is running in Automatic or Manual mode When in Monitor mode no changes will be made to the COF PF of the control instrumen...

Страница 19: ...COF PF Adj C Press the down arrow key once to highlight the On Delay field Enter a 10 5 and press Enter Press the down arrow key once to highlight the Off Delay field Enter a 5 and press Enter Press t...

Страница 20: ...inue If the value is more than 20 out it will be BAD and the calibration will not be allowed to proceed If this occurs check to make sure that you are using the correct gas and that there is adequate...

Страница 21: ...with the position number are Position Rate Position Rate 0 1200 9 76800 1 2400 10 115200 2 4800 11 230400 3 9600 12 460800 4 14400 13 921600 5 19200 6 28800 7 38400 8 57600 To change the baud rate pre...

Страница 22: ...the circular arrow key to cycle through the options or press the option s position number and press Enter The Furn TC Inst Addr is the address of the Furn TC Inst To change the address enter the new v...

Страница 23: ...out it will be BAD and the calibration will not be allowed to proceed If this occurs check to make sure that you are using the correct gas and that there is adequate flow If it is still not operating...

Страница 24: ...ions address The instrument does not have DHCP therefore a fixed IP address must be assigned General Information Menu Page 28 This menu page will display some general information about the PGA This sc...

Страница 25: ...n Gas Part Number 13084 Bowl Filter Assembly Including Element Part Number 37048 Bowl Filter Element Media Replacement 10 Pack Part Number 31027 Flexible Sample Tubing Assembly with Filter Part Number...

Страница 26: ...7 2041 C Added MCO section in Revision History Updated logo on title page Updated parts list Changed picture sizes Added functionality for menus 8 14 15 20 28 Updated menu 6 to include help text Updat...

Отзывы: