SSi Manual 4559 

Page 12 of 26 

3-Gas Analyzer PGA3500 

Revision Level - D 

large changes in CO Factor or Process Factor should be verified by shim stock analysis or 

other means to confirm the significance of the change. 

 

Combustion Display – Menu Page 3 

The combustion display shows the current 

reading in % excess oxygen.  Also shown is 

the relative flow rate of the sample by a fuel 

gauge on the left-hand side of the screen.  

The sample pump may be turned on or off 

by pressing the circular arrow key. 

This screen is not intended to measure 

oxygen as accurately as a probe.  It is 

intended only to serve as a method of 

checking to see if there is oxygen 

contamination in the sample.  This sensor 

measures oxygen to a tenth of a percent, 

where a probe will provide accuracy in the 

parts per million. 

 

Pump Control - Menu Page 4 

The pump control page is used to set the pump 

mode. The sample pump mode can be changed from 

OFF to ON to AUTO by pressing the circular arrow 

key.  Auto mode will turn the pump on when the flow 

is below approximately 1.5 SCFH and off when it is 

above approximately 1.5 SCFH.  Also shown is the 

relative flow rate of the sample by a fuel gauge on 

the left-hand side of the screen.  There is also a 

traditional flow meter located in the lid of the PGA.  

Although the flow indicator on the screen has been 

calibrated at the factory, the most accurate flow 

measurements should be taken with the flow meter 

inside the lid. 

 

Set Display Values – Menu Page 5 

 

This page is used to adjust the display values.  

The backlight brightness and contrast are 

factory set to 51%, which should be 

appropriate for most viewing conditions.  Their 

values can range from 0 to 100% if desired.  

Also shown on this page is the Backlight ON 

Time.  This is a power saving feature that will 

enhance battery life by turning the backlight 

off if no buttons are pressed within a pre-

determined number of minutes.  When it turns 

Содержание PGA 3500

Страница 1: ...ns before operating this equipment Super Systems Inc is not responsible for damages incurred due to a failure to comply with these instructions If at any time there are questions regarding the proper...

Страница 2: ...Display Values Menu Page 5 12 Help Page Menu Page 6 13 Calibration Dates and Run Times Menu Page 7 13 Language Lengua Menu Page 8 13 Battery Status Menu Page 9 14 About Sign On Menu Page 10 14 Revisi...

Страница 3: ...ng reshipment Super Systems will not be responsible for damage caused by improper packaging Moisture and Condensation Depending on the composition of the gas being sampled it is possible that condensa...

Страница 4: ...ent Measuring the level of CO in the carrier gas allows correction of the carbon reading at the furnace Heat Treat Furnaces Conventional Endo Gas Furnace atmosphere carbon potential C can be verified...

Страница 5: ...e CO and CH4 sensors have been calibrated to be most accurate within the ranges as shown above however both sensors are capable of measuring gas concentrations of up to 100 00 Sampling method Extracti...

Страница 6: ...uously pressed toggles between the main page and the menu list On most pages this key can be used to turn the pump on and off without using Screen 4 On some screens it is used to change a selection or...

Страница 7: ...ure smooth operation When new the elements in both of these filters are a yellow white color Both are oriented in the sample stream in a manner that causes any contaminants to collect on the outside o...

Страница 8: ...SET DISPLAY VALUES 6 HELP 7 CALIBRATION DATES 8 LANGUAGE LENGUA 9 BATTERY STATUS 10 ABOUT SIGN ON 11 REVISION DISPLAY 12 LOGGED DATA START DATE 13 LOGGED DATA DISPLAY 14 SESSION 15 SAMPLING PARAMETERS...

Страница 9: ...and also displays the internal temperature IT of the instrument This internal temperature should never exceed 50 Note about Menu Numbers Each menu screen has a unique number that will be displayed in...

Страница 10: ...area up and down the following data should be entered FC TC This is the furnace thermocouple value or the furnace temperature PB MV This is the millivoltage from the carbon probe PB TC This is the pro...

Страница 11: ...s not to calculate the probe carbon but to enable the PGA to compute the suggested COF PF Without knowing the current instrument values it cannot compute the suggested values What is a CO Factor or a...

Страница 12: ...set the pump mode The sample pump mode can be changed from OFF to ON to AUTO by pressing the circular arrow key Auto mode will turn the pump on when the flow is below approximately 1 5 SCFH and off wh...

Страница 13: ...hlight the desired location by using the up and down arrow keys Then enter the number using the numeric keypad and pressing Enter Pressing the ESC button once will clear any typed data that is on the...

Страница 14: ...recharge the battery but it will take a few days to accomplish this By leaving the instrument plugged in and periodically checking the battery voltage to see if it is increasing you will be able to de...

Страница 15: ...date Logged Data File Menu Page 13 Each minute there are 15 points of data collected The time of each data point is shown along the left hand side of the screen The arrow buttons will scroll backwards...

Страница 16: ...art a session on the PGA A session will allow the user to identify the operation of the PGA and break all of the logged data into manageable sections Sessions can be viewed on a computer with the XGA...

Страница 17: ...ocations can all be modified and added through the XGA Viewer software Once a session has started the second page of the session menu will be displayed From this page the user can add a note to the se...

Страница 18: ...range from 0 to 999 The IR Mode is a selection to determine if the instrument is running in Automatic or Manual mode When in Monitor mode no changes will be made to the COF PF of the control instrumen...

Страница 19: ...COF PF Adj C Press the down arrow key once to highlight the On Delay field Enter a 10 5 and press Enter Press the down arrow key once to highlight the Off Delay field Enter a 5 and press Enter Press t...

Страница 20: ...inue If the value is more than 20 out it will be BAD and the calibration will not be allowed to proceed If this occurs check to make sure that you are using the correct gas and that there is adequate...

Страница 21: ...with the position number are Position Rate Position Rate 0 1200 9 76800 1 2400 10 115200 2 4800 11 230400 3 9600 12 460800 4 14400 13 921600 5 19200 6 28800 7 38400 8 57600 To change the baud rate pre...

Страница 22: ...the circular arrow key to cycle through the options or press the option s position number and press Enter The Furn TC Inst Addr is the address of the Furn TC Inst To change the address enter the new v...

Страница 23: ...out it will be BAD and the calibration will not be allowed to proceed If this occurs check to make sure that you are using the correct gas and that there is adequate flow If it is still not operating...

Страница 24: ...ions address The instrument does not have DHCP therefore a fixed IP address must be assigned General Information Menu Page 28 This menu page will display some general information about the PGA This sc...

Страница 25: ...n Gas Part Number 13084 Bowl Filter Assembly Including Element Part Number 37048 Bowl Filter Element Media Replacement 10 Pack Part Number 31027 Flexible Sample Tubing Assembly with Filter Part Number...

Страница 26: ...7 2041 C Added MCO section in Revision History Updated logo on title page Updated parts list Changed picture sizes Added functionality for menus 8 14 15 20 28 Updated menu 6 to include help text Updat...

Отзывы: