The display shows:
Pressing the [Left (Back)] button takes you back
to [MAIN MENU] .
Select [Calibration] using the [Up] and [Down]
buttons .
The display shows:
Confirm your selection by pressing [OK] or
[Right (Next)] .
The display shows:
Remove the barcode strip for calibration from
the pack (included in the scope of delivery of
RQflex
®
20) .
Insert the barcode all the way into the barcode
reader in the direction of the arrow in a smooth
motion from left to right, and then remove it
again .
Make sure that no test strip ist inserted in the
test strip adapter, then press [OK] .
The display shows:
Insert the calibration strip of the recalibration
set (included in the scope of delivery of
RQflex
®
20) into the strip adapter with the
beveled edge outwards and upwards and press
the [START measurement] button .
The calibration is carried out .
The display shows:
Pressing the [OK] button leads you forward to
[Calibration Results] (see „Calibration Results“,
5 .1 .2) .
Abort:
For abort press the [Menu] button .