23
22
2 .6 .10 Setting the memory
capacity warning (ON or OFF)
Select [Memory Capacity Warning] using the
[Up] and [Down] buttons .
The display shows:
Confirm your selection by pressing [OK] or
[Right (Next)] .
The display shows:
Select the desired memory space warning mode
(On or Off) using the [Up] and [Down] buttons .
Confirm your selection by pressing [OK] or
[Right (Next)] and return to [SETTINGS] .
Press the [Left (Back)] button to return to
[MAIN MENU] .
Abort:
Press the [Left (Back)] button to abort if
desired, and return to [SETTINGS] .
2.7
Initial calibration
The instrument needs to be calibrated before the
first measurement is performed . This is required
to set the correct baseline for the reflectome-
tric optics . For this purpose, a Reflectoquant
®
Recalibration Set is delivered with the instru-
ment . Please use the barcode and the pla-
stic test strip for calibration (for details of
the Recalibration Set see section 5, "Quality
assurance") .
Keep the instrument at room temperature for at
least 30 minutes .
Switch on the reflectometer by pressing the
[Power On/Off]
button .
The unit runs an electronic self-check test .
The display then shows the main menu:
Select [Quality Assurance] in the [MAIN
MENU] using the [Up] and [Down] buttons .
The display shows:
Confirm your selection by pressing [OK] or
[Right (Next)] .