51
Safety systems: Independent hardware over-pressure circuit and redundant software
overpressure algorithm to limit cuff pressure to less than 330 mmHg. Independent
hardware timing circuit and redundant software timer algorithm to limit the duration
of a blood pressure cycle to less than 180 seconds.
Dimensions: Length = 5.5 inches, Height = 11.5 inches, Width = 3.8 inches; Length =14.0 cm,
Height = 29.2 cm, Width = 9.7 cm
Standards: UL60601-1, CAN/CSA C22.2 601-1
IEC 60601-1, IEC 60601-1-2 (EMC), IEC 60601-1-4, ISO 9919, AAMI SP10:2002,
ASTM E 1112, EN 12470-3
Meets EN-1060-1, Specification for non-invasive sphygmomanometers – Part 1:
General requirements and EN 1060-3, Non-invasive Sphygmomanometers – Part 3.
Supplementary Requirements for Electro-Mechanical BP Measuring Systems”
Classification: Protection against electric shock: Class II (for non-battery version), Internally
Powered Equipment (for battery version); Applied parts: Type BF; Mode of operation:
Continuous
IP index: IPX0
Compliance
Compliance
Compliance
Compliance
SunTech Medical, Ltd.
Oakfield Industrial Estate
Eynsham, Oxfordshire OX29 4TS
UK
Tel: +44. 1865.884.234
Fax: +44. 1865.884.235
Safety Requirements
Clinical grade BP measurement accuracy defined by fully meeting the requirements of:
•
AAMI SP-10 2002
•
EN 1060-4