TROUBLESHOOTING
TROUBLESHOOTING
TROUBLESHOOTING
TROUBLESHOOTING
The troubleshooting chart provides pointers on diagnosing issues associated with error or
status codes.
Troubleshooting - Blood Pressure Module
Problem: Wrong size cuff, Misplaced cuff, or Blocked brachial artery
Solutions:
1.
Check that the cuff is in the correct position.
2.
Check that the cuff is properly tightened.
3.
Check that there is no excessive clothing between the arm and the cuff.
4.
Check that the cuff applied is of the correct size.
5.
The patient may have been moving too much.
6.
Take another BP reading.
Problem: Too much patient or environment motion or conditions causing tremors
Solutions:
1.
Check that the cuff is in the correct position.
2.
The patient may have been moving too much.
3.
Take another BP reading.
Problem: Air leak, Loose cuff, or Blocked or pinched hose
Solutions:
1.
Check that the hose has no sharp bends or is pinched.
2.
Check that the patient is not lying on the cuff.
3.
Check that the cuff is in the correct position.
4.
Check that the hose is connected to the system and the cuff.
5.
Check that the cuff is properly tightened.
6.
Check that the correct size cuff is being applied.
7.
Check that the cuff is not leaking air.
8.
Check that the hose connections are not damaged or loose.
9.
Take another BP reading.