Sunrise Medical Quickie Q200 R Скачать руководство пользователя страница 6

Quickie

®

 Q200

®

 R General Manual

253470 Rev. B

6

2.0 SAFETY

2.5 Safety: Choking hazard

DANGER!

This mobility aid uses small parts which under certain circumstances may 
present a choking hazard to young children.

2.6 Safety: Using a (vehicle mounted) wheelchair lift

Wheelchair lifts are used in vans, buses, and buildings to help you move from 
one level to another.

DANGER!

1. 

Ensure that the user and all caregivers fully understand the lift 
manufacturer’s instructions for using the passenger lift.

2. 

Never exceed the lift manufacturer’s recommended safe working load 
and load distribution guidance.

3. 

Always turn off  all power when you are on the lift. If you fail to do so, 
you may touch the joystick by accident and cause your chair to drive off  
the platform. Be aware that a roll-stop at the end of the platform may 
not prevent this.

4. 

Always position the user securely in the chair to help avoid falls while on 
the lift.

5. 

Always ensure the chair is in drive mode when using passenger lift 
(wheels locked, not in freewheel mode).

2.7 Safety: Accessory Trays

Safety: Vent tray

DANGER!

Using a vent tray will aff ect the stability and overall weight of your wheelchair.
1. 

Ensure that the batteries and the ventilator are securely fastened to the 
tray before use.

2. 

Familiarize yourself with the increased size of the wheelchair before 
driving to prevent potential collisions.

3. 

When activating the recline backrest ensure that the area around the 
vent tray is clear.

4. 

Be aware that the vent tray assembly can cause a fi nger trap hazard when 
the backrest is reclined.

5. 

The vent tray is not designed to hold anything other than the vent unit 
and its batteries. It is not designed to hold an oxygen bottle etc. Use of 
these items may cause your wheelchair to become unstable.

6. 

The vent tray must be installed and maintained by a Sunrise Medical 
authorized dealer.

2.8 Safety: Lifting the wheelchair

WARNING!

Do not lift this seating system by any parts that are removable. Doing so may 
result in damage to the seating system or injury to the user.

2.9 Safety: Stability

WARNING!

1. 

Removing the batteries may cause the chair to become unstable causing it 
to tip backwards.

2. 

Only remove the batteries while the chair is on the ground.

3. 

Do not remove the batteries while the chair is on a lift table or on any 
other elevated surface.

Содержание Quickie Q200 R

Страница 1: ...Medical com Ingrese a la p gina del producto espec fi co para descargar el manual o comun quese con el proveedor autorizado de Sunrise Medical Au Fournisseur Ce manuel doit tre remis l utilisateur de...

Страница 2: ...ealer ______________________________________________________ Address _____________________________________________________ Telephone __________________________________________________ Serial _________...

Страница 3: ...4 6 Check after setup and or adjustment of wheelchair 9 5 0 USING THE WHEELCHAIR 10 5 1 Checking wheelchair before use 10 5 2 Making a transfer 10 5 3 Driving the wheelchair 10 5 4 Curves 11 5 5 Brea...

Страница 4: ...seating system The user manual for the controller The user manual for the battery charger When necessary this general user manual refers to other manuals as shown below Seating system Refers to the u...

Страница 5: ...e back and the armrests can become hot if they have been exposed to full sunlight for too long This may cause burns or allergic reactions to the skin 2 3 Safety Moving parts DANGER A wheelchair has mo...

Страница 6: ...Vent tray DANGER Using a vent tray will affect the stability and overall weight of your wheelchair 1 Ensure that the batteries and the ventilator are securely fastened to the tray before use 2 Familia...

Страница 7: ...it includes both the user and any weight of accessories fitted to the wheelchair is marked on the serial number label which is affixed to the chair The overall maximum user weight shall not exceed Sta...

Страница 8: ...r and or damaged or broken some of the components Wheelchairs involved in sudden stops should be inspected for possible failures in frame and or components Frame damage may be represented by but not l...

Страница 9: ...Adjusting the center of gravity on a powered wheelchair is both physically demanding and safety critical Sunrise Medical strongly recommends that this adjustment is only performed by a Sunrise Medica...

Страница 10: ...he normal capacity At temperatures below 23 F 5 C this will be roughly 50 This will reduce your range of action 5 2 Making a transfer Sunrise Medical recommends that you consult your health care provi...

Страница 11: ...adients in standard configuration 7 5 13 DANGER 1 Elevating tilting and leg rest adjustments can affect the wheelchair center of gravity Under certain conditions this might cause instability that coul...

Страница 12: ...s far as possible keep the joystick in the straight forward position The maximum obstacle or curb climbing height is 2 3 in 60 mm for Group 3 The approach speed and process can vary depending on your...

Страница 13: ...heelchair is in push 3 Never set the switch to the push position on a slope When the freewheel switch is set to push the automatic parking brake is deactivated This makes it possible for the wheelchai...

Страница 14: ...5 Do not attempt to operate any power seating option while this seating system is in motion 6 Do not attempt to operate the tilt power seating options around children 7 Through programming it is poss...

Страница 15: ...ld a charge 4 The battery terminals need to be checked regularly for signs of corrosion If any corrosion is apparent clean the terminals completely and apply petroleum jelly or silicone grease to the...

Страница 16: ...r place the charger directly above the battery being charged gases from the battery will corrode and damage the charger 6 Never charge a frozen battery A fully charged battery will rarely freeze but t...

Страница 17: ...the battery shroud To disconnect the batteries 1 Remove the front battery cover by lifting upwards on the handle at the front of the cover Fig 6 2 2 Undo the two GREY plugs that connect the two batte...

Страница 18: ...access and maneuverability to a forward facing position Securing the wheelchair with a strap restraint system This wheelchair has been successfully crash tested in accordance to ISO 7176 19 2008 using...

Страница 19: ...cupant Restraint Systems For Use in MotorVehicles Do not use WTORS designed to rely on the wheelchair structure to transfer occupant restraint loads to the vehicle 12 The wheelchair has been dynamical...

Страница 20: ...packed and labeled so they do not get lost during loading and unloading 4 The wheelchair may be transported by road rail sea or air and the batteries conform to IATA regulations 5 Before you travel pl...

Страница 21: ...locking nuts or nuts studs secured with a thread locking solution ensure that a suitable thread locking solution is reapplied to the fastener Daily checks Perform the daily check routine before drivi...

Страница 22: ...eplacing self locking nuts or nuts studs secured with a thread locking solution ensure that a suitable thread locking solution is reapplied to the fastener 6 Check all Velcro fastening straps for corr...

Страница 23: ...unctions properly it is essential that the tires are kept at the correct pressure The correct tire pressure is indicated on the side of the tire Tires that are too soft will have a negative effect on...

Страница 24: ...ng re used Prior to the wheelchair being re used it must be carefully prepared All surfaces which come into contact with the user must be treated with a disinfection spray To do this you must use a su...

Страница 25: ...chair off and then back on again 3 Check whether the battery plugs are all securely in place 4 Check whether the freewheel switch is in the DRIVE position 5 Check the position of the speed regulator 6...

Страница 26: ...4 2014 Class B Electrically powered wheelchairs scooters and their chargers Requirements and test methods 300 lbs 136 kg ISO 7176 8 2014 Requirements and test methods for impact static and fatigue str...

Страница 27: ...turn or repair of covered parts Contact Customer Service for a Returned Material Authorization RMA number and information as to where to return the product Upon contacting Customer Service please prov...

Страница 28: ...Customer Service 800 333 4000 In Canada 800 263 3390 Sunrise Medical 2842 Business Park Ave Fresno CA 93727 USA 2021 Sunrise Medical US LLC 10 2021 P N 253470 Rev B www SunriseMedical com...