![Sunrise Medical Quickie Pulse 6 BC Скачать руководство пользователя страница 61](http://html1.mh-extra.com/html/sunrise-medical/quickie-pulse-6-bc/quickie-pulse-6-bc_service-manual_1399832061.webp)
PULSE 6 SERVICE MANUAL
PAGE 7.14
2009
SUNRISE MEDICAL
Section 7
4.
Reconnect the 6-pin connector on the rear
of the tilt and partially tilt the seating system back
to expose the actuator. Disconnect the 6-pin
connector
5.
Locate the tilt actuator under the seat (
fi
g.
7.41).
Note: Complete this step only if the micro-switch assembly is to be replaced separately.
6.
Locate the micro-switch assembly at the rear
of the tilt (
fi
g. 7.42). Note the connections leading
to the micro-switch. Separate the wires leading
to the micro-switch. Reverse the procedures to
replace it and retest as necessary.
Warning: The tilt system is free to fall during
the next step. Secure the seating system prior
to removing the tilt actuator.
7.
Use two 13 mm wrenches and remove
the nuts at both ends of the actuator (
fi
g. 7.43).
Remove both pins from the tilt actuator.
Removal Procedures (cont.)
fig. 7.42
fig. 7.41