Over-current
This fault occurs when the limits of an actuator circuit is exceeded.
Perform Tilt Will Not Operate troubleshooting section of this manual to test
the end of travel limit switches.
Overtemp. (Acts)
This error indicates that the Intelligent Seating Module (ISM) has become
excessive warm. Note: An ISM is only required when there are more than
2 actuators and may not be mounted on this power chair. Allow the unit to
cool. If the error repeats, the actuator may be defective or over loaded.
Overtemp. (Lamps)
Note: This fault can only occur if an Intelligent Seating Module is used on
this power chair and connected to external lights.
DIME Error
This error indicated that one or more of the modules are not compatible.
Remove the last module installed and retest. Recycle the power. Contact
Sunrise Technical Service for assistance.
Memory Error
Recycle the power. Verify that the cables are correctly mated. If a module
has been replaced recently, suspect that module may be defective.
Contact Sunrise Technical Service for assistance.
PM Memory Error
Recycle the power. Verify that the cables are correctly mated. If a module
has been replaced recently, suspect that module may be defective.
Contact Sunrise Technical Service for assistance.
Bad Cable
Inspect and replace the defective cable(s).
Bad Settings
Verify that the programming agrees with the installed equipment. If
all settings are correct, the control module may be defective. Contact
Sunrise Technical Service for assistance.
Module Error
Recycle the power. Verify that the cables are correctly mated. If a module
has been replaced recently, suspect that module may be defective.
Contact Sunrise Technical Service for assistance.
System Error
Recycle the power. Verify that the cables are correctly mated. If a module
has been replaced recently, suspect that module may be defective.
Contact Sunrise Technical Service for assistance.
SID Detached
The Omni has detected that a specialty control has become disconnected.
Recheck all cables. If the error is still present, replace the specialty
control.
User Switch Detached
Indicates that a user switch has become disconnected. Reconnect the
switch.
Gone to Sleep
The predetermined sleep time has been exceeded due to inactivity by the
user.
Charging
This indication is present when the battery charger is connected. There
may also be an error in the control module. Contact Sunrise Technical
Service for assistance.