Sunnex PF Series MRI Service & Installation Manual
25
Guidance and Manufacturer’s Declaration
—
Emissions Equipment
and Systems that are NOT Life-Supporting
The PF MRI Light (PF 28016-M) is intended for use in the electromagnetic
environment specified below. The customer or user of the PF MRI Light
(PF 28016-M)
should ensure that it is used in such an environment.
Immunity
Test
IEC
60601
Test
Level
Complia
nce
Level
Electromagnetic Environment
—
Guidance
Conducted
RF
IEC 61 000-
4-6
3 Vrms
from
150 kHz
to
80 MHz
V1 = 3
Vrms
Portable and mobile RF
communications equipment should
be separated from the PF MRI Light
(PF 28016-M)
by no less than the
recommended separation distances
calculated/listed below:
D
=
(3.5/V1)
P
Radiated
RF
IEC 61000-
4-3
3 V/m
80 MHz
to
2.5 GHz
E1 =
3V/m
D
= (3.5/ E1)/
P
80 to 800 MHz
D
= (7 / E1)/
P
800 MHz to 2.5 GHz
Where P is the maximum power
rating in watts and D is the
recommended separation distance in
meters.
Field strengths from fixed
transmitters, as determined by an
electromagnetic site survey, should
be less that the compliance levels
(V1 and E1).
Interference may occur in the vicinity
of equipment containing a transmitter.