APPENDIX

APPENDIX

Guidance and manufacturer’s declaration – electromagnetic emissions

The device is intended for use in the electromagnetic environments listed below, and should only be 

used in such environments:

Emission test

Compliance

Electromagnetic environment – guidance

RF emissions

CISPR 11

Group 1

RF energy is used only to maintain device’s 

operation. Therefore, its RF emissions are so low 

that it’s not likely to cause any interference in 

nearby electronic equipment. 

RF emissions

CISPR 11

Class B

The device is suitable for use in all establishments, 

including domestic establishments, and those 

directly connected to the public low-voltage power 

supply network that supplies buildings used for 

domestic purposes.

Harmonic emissions IEC 

61000-3-2

Class A

Voltage fluctuations / flicker 

emissions IEC 61000-3-3 

Complies

Guidance and manufacturer’s declaration – electromagnetic immunity

The device is intended for use in the electromagnetic environments listed below, and should only be 

used in such environments:

Immunity 

test

IEC 60601 

test level

Compliance level

Electromagnetic 

environment - guidance

Electrostatic 

discharge (ESD)

IEC 61000-4-2

±8kV contact discharge

±15kV air discharge

±8kV contact 

discharge

±15kV air discharge

In the case of air discharge 

testing, the climatic conditions 

shall be within the following 

ranges: Ambient Temperature: 

15°C - 35°C, Relative 

Humidity: 30% - 60%. 

Power frequency

(50 or 60Hz)

magnetic eld

IEC 61000-4-8

30A/m

50 or 60Hz

30A/m

50 or 60Hz

Power frequency magnetic 

elds should be at levels 

characteristic of a typical 

location in a typical commercial 

or hospital environment.

Electrical fast

transient / burst

IEC 61000-4-4

± 2 kV for power

supply lines

± 1 kV for input / 

output lines

± 2 kV for power

supply lines

± 1 kV for input / 

output lines

Mains power quality 

should be that of a typical 

commercial or hospital 

environment.

Surge

IEC 61000-4-5

AC Power port

±1 KV Line to Line

AC Power port

±1 KV Line to Line

Mains power quality 

should be that of a typical 

commercial or hospital 

environment.

interruptions and

voltage variations

on power supply

input lines

IEC 61000-4-11

0% UT; 0.5 cycle

At 0°,45°,90°, 

135°,180°,225°,

270°and 315°.

0% UT; 1 cycles

70% UT; 25 / 30 cycles

0% UT; 250 / 300 cycle

0% UT; 0.5 cycle

At 0°,45°,90°, 

135°,180°,225°,

270°and 315°.

0% UT; 1 cycles

70% UT; 25 / 30 cycles

0% UT; 250 /300 cycle

Mains power quality 

should be that of a typical 

commercial or hospital 

environment. If the user 

of the device requires 

continued operation during 

power mains interruptions, 

it is recommended that the 

device be powered from an 

uninterruptible power supply 

or a battery.

NOTE:

 U

T

 is the a.c. mains voltage prior to application of the test level.

 

Содержание 16994

Страница 1: ...1 Upper Arm Blood Pressure Monitor with Voice Broadcast Technology Instruction Manual Model HL858A2 LT Item 16994 Manual Version 001 Issue Date XX ...

Страница 2: ...curate idea of your blood pressure MEASUREMENT METHOD HL858A2 LTAutomatic Upper Arm Blood Pressure Monitor measures blood pressure and heart rate by oscillometric method meaning the fluctuations in pressure are measured Once the cuff is wrapped around your upper arm just turn on the monitor and inflation automatically starts The inflation of the cuff creates pressure around the arteries inside upp...

Страница 3: ...osis diabetes liver disease kidney disease severe hypertension peripheral circulation please consult your healthcare professional before using the device Blood pressure measurements taken with this device are equivalent to those obtained by a trained observer using the cuff stethoscope auscultation method and are within the accuracy limits prescribed by the Standard of EN 1060 4 If the cuff is wor...

Страница 4: ...perly BP Category Indicator The arrowhead points out the specific BP Category that your measurement reading fits in Average of Last 3 Measurements ThissymbolappearswhenLCDdisplaysaverage value of last3readings IRREGULAR HEARTBEAT DETECTOR The symbol will appear on screen indicating a certain heartbeat irregularity was detected during measurement The heartbeat rhythm that is more than or less than ...

Страница 5: ... Insert 4 new AAA 1 5 V LR03 alkaline batteries into the compartment as shown Make sure the polazrities and align with the symbols inside the compartment 3 Replace the batteries when The low battery icon appears The display is dim or does not turn on Do not mix new and used batteries Do not mix alkaline standard carbon zinc or rechargeable cadmium batteries Such action may shorten the battery life...

Страница 6: ...uff inanyway MEMORY FUNCTION 1 Themonitorfeaturesmemorystoragewithdateandtimestampsfortwo 2 users saving 120recordsperuser 240total 2 Aftereachmeasurement thesystolicanddiastolicpressure heartrateandBPCategory Indicator IrregularHeartbeatDetectorandExcessivebodymotiondetector if any withthe dateandtimewillbeautomaticallystored Eachnewreadingwillreplacetheoldest TAKING THE MEASUREMENT 1 With the cu...

Страница 7: ...ol The cuff was wrapped incorrectly for example too loosely or too tightly Please reference applying the Cuff section to wrap the cuff correctly Note If EP appears on the display just return the device to your local distributor or importer RECALLING DATA 1 Press buttontoselectUser1or2 2 PressM buttontoenterMemoryMode If thereisnodatastored nothing exceptmonth date andtime willappearonthedisplay If...

Страница 8: ...hthelimitsforaClassBdigital device pursuanttoPart15of theFCCRules Theselimitsaredesignedtoprovidereasonable protectionagainstharmfulinterferenceinaresidentialinstallation Thisequipmentgenerates usesandcanradiateradiofrequencyenergyand if notinstalledandusedinaccordancewith theinstructions maycauseharmfulinterferencetoradiocommunications However there isnoguaranteethatinterferencewillnotoccurinapar...

Страница 9: ...15kV air discharge In the case of air discharge testing the climatic conditions shall be within the following ranges AmbientTemperature 15 C 35 C Relative Humidity 30 60 Power frequency 50 or 60Hz magnetic eld IEC 61000 4 8 30A m 50 or 60Hz 30A m 50 or 60Hz Powerfrequency magnetic eldsshouldbeatlevels characteristicof atypical locationinatypicalcommercial orhospitalenvironment Electrical fast tran...

Страница 10: ...ed by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80MHz and 800MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures object...

Страница 11: ...399 8 30 am 5 00 pm EST Monday Friday or contact us at info stareliteinc com Our customer service will be happy to assist you 2021 Sunbeam Products Inc All rights reserved Distributed by Star Elite Inc Montreal Canada H3B 3X9 SE005 080621 Upper Arm Blood Pressure Monitor Model HL858A2 LT Printed in China ...

Отзывы: