background image

If necessary, carry out a monthly visual inspection. To clean the cell:

Remove the cell from its support (after turning off the filtration system and closing off the necessary valves).
Place the cell for no more than 10 minutes in 15% hydrochloric acid (1,5 l of acid for each 8,5 l of water).
Once the incrustations have softened remove with a hose to complete cleaning the cell.

DO NOT USE METALIC OR SHARP OBJECTS TO REMOVE INCRUSTATIONS. Scratching the edges or surface of the cell will make it 
vulnerable to chemicals, deteriorate the cell and cancel the guarantee.

5

SYSTEMS WITH redoX CONTROL

6

MAINTENANCE

During the first 10-15 days your pool system will require more attention and the following care:

Make sure the pH remains on the ideal level (7,2 - 7,5). If the pH is unusually unstable and uses a lot of acid check the alkalinity 

(recommended levels between 80 y 120 ppm). 
The pool must be vacuumed and the skimmers cleaned whenever necessary to ensure perfect water conditions.

REMEMBER that the system requires a certain amount of time to adapt to your swimming pool and will require additional chemicals during the first 3-5 days.

The pool must be vacuumed as usual and the skimmers emptied whenever necessary.

FILTER BACKWASHING: The system requires only occasional filter cleaning; once every 20 days should be sufficient (providing  

the filter pressure does not exceed 1 bar, in which case a filter cleaning may be necessary).

VERY IMPORTANT: Make sure the cell is off while cleaning the filter. If the system controls the filtration pump, use the option “filter cleaning” of the programmed 

filtration mode. See section 4.3 – Filtration (Filter Cleaning).

ADDING NEW WATER: Always through the skimmers so that the new water passes through the Oxilife before entering the pool. 

Remember to add the necessary salt (1,5-2,5 gr) per added liter of water.

In winter changing the pool water is not recommendable. We recommend that the system runs 2-3 times per week (2-3 hours per day).

DOSING PUMPS: Check regularly to ensure that the container contains liquid to prevent the dosing pump of running dry. The dosing 

pump requires maintenance (SEE INSTRUCTIONS ON BOX).

pH PROBES / redoX / CONDUCTIVITY: Probes must be cleaned whenever necessary (check every 5-6 months). To clean the probe 

insert in distilled water (clear liquid). After each cleaning the probes must be calibrated. Also: the probes should never dry out and 

must be kept wet if stored (when emptying the pool for winterizing, make sure to store the measuring head in water). 

80 - 120 ppm
1.500 - 2.500 ppm

30 - 50 ppm

Visual inspection to 
detect incrustations.

1,0 - 2,0 ppm
7,2 - 7,5

FREE CHLORINE:

pH:

TOTAL ALKALINITY (TAC) pH:

SALT CONCENTRATION:

CYANURIC ACID:

TITANIUM CELL:

1

1

1

2

2

2

3

3

4

5

6

The redoX value advises us of the oxidation/reduction potential and is used to determine the level of water sterilization.  
The parameters or setpoints are the minimum/maximum accepted redoX levels before the titanium cell is connected/disconnected. 
Adjusting the ideal redoX level (setpoint) is the last step in the Oxilife start up sequence. To find the optimum redoX levels for your  
pool follow these steps:

Connect the pool filtration system (the salt in the pool must be adequately dissolved).
Add chlorine to the pool till a level of 1-1,5 ppm is achieved (approx. 1-1,5 gr/ m

3

 of water). pH levels should be between 7,2 - 7,5.

After 30 min. test the free chlorine levels in the pool (manual test kit DPD1) if the free chlorine level is between 0,8 - 1,0 ppm.  

Look at the redoX screen and memorize this level as the setpoint to CONNECT / DISCONNECT the low salinity electrolysis cell.
The next day check free chlorine levels (manual test kit DPD1) and redoX. Raise / lower setpoint if necessary.
Remember to check the redoX set-point every 2-3 month and/or if the water parameters change (pH / temperature / conductivity).

1
2
3

4
5

First days of maintenance

Cleaning the titanium cell  

General maintenance

Fortnightly checks

Monthly checks

Содержание oxilife

Страница 1: ...for advanced functions If you are an authorized installer please ask your provider for the installers service manual Connection WIFI RS module Device control SLAVE Digital inputs outputs Connection display Electrical connections of the electronic box Chlorine probe detector rotameter Cover FL1 Conductivity probe Temperature probe pH redoX Free Chlorine Low salinity electrolysis cell RCA flow detec...

Страница 2: ...lt concentrations from 2 5 g l to 50 g l s In pools with strong insolation it s necessary to add 40 gr m3 of stabiliser isocyanuric acid Water adjustments Adding salt to the water 4 5 6 3 INITIAL WATER ADJUSTMENTS hydrolysis measures pH 7 5 OFF OFF 7 0 7 2 Rx mV 700 OFF FL1 700 man off man off 12 30 25º 100 1 2 3 4 Cover Pol 1 Low 4 FUNCTIONING OF THE SYSTEM Current time Hydrolysis Intensity of th...

Страница 3: ... actual temperature use the plus minus and up down keys Set to the actual temperature of the probe and press OK 2 9 Calibration of the redoX probe Recommended every 2 months during usage season 2 10 Calibration with buffer buffer liquid 465 mV Follow the instructions which appear on the displays fig 2 11 You can access the display Filter cleaning from any filtration mode Once selected this functio...

Страница 4: ...ting of preferred language 6 5 Setting of day and current time 6 7 Setting of the intensity of the display lighting 0 100 and programming its ON and OFF time 4 6 Settings 6 12 6 14 6 13 6 1 6 3 6 4 6 2 6 5 6 8 6 9 6 6 6 11 6 7 6 10 2 x Clima OFF The heating only works within the set filtration periods Clima ON Keeps the filtration working when the filtration period is finished if the water tempera...

Страница 5: ...ol water is not recommendable We recommend that the system runs 2 3 times per week 2 3 hours per day DOSING PUMPS Check regularly to ensure that the container contains liquid to prevent the dosing pump of running dry The dosing pump requires maintenance SEE INSTRUCTIONS ON BOX pH PROBES redoX CONDUCTIVITY Probes must be cleaned whenever necessary check every 5 6 months To clean the probe insert in...

Страница 6: ...section 3 Initial water adjustments s Check for incrustations on cell s See section 7 Electrolysis does not reach maximum intensity Electrolysis display shows LOW s The water is excessively hard and it is unbalanced s Balance the water and check the cell proceeding to clean it if necessary s Put 1 small bag of flocculant in the skimmer and recirculate 24 hours White flakes in the water s Check flo...

Отзывы: