OPERATING MANUAL FOR SELECT EDI 60
L300180 Iss:A
32
LHS PACK NOT
FITTED
PT media pack not fitted or not fully
engaged.
Check pack is fully engaged and
retainer pin locked. (ref sec 7.1)
RHS PACK NOT
FITTED
Flushing pack not fitted or not fully
engaged.
Check Pack is fully engaged and
retainer pin located. (Ref Sec 7.1)
FIT BOTH PACKS
PT media pack or Flushing pack
not fitted or both not fully engaged.
Check pack/s are fully engaged and
retainer pins located. (Ref Sec 7.1)
TIME/DATE NOT SET
On Start Up, time and date may not
be set.
Main PCB battery needs replacing.
Select Set Time and Date function
from settings menu and set.
Replace lithium PCB battery (Ref
Sec. 7.4)
POOR WATER
QUALITY
RO membranes fouled
Check membrane performance,
clean if required.
Contact supplier if new pack does
not improve water quality.
TANK LOW LEVEL
Water in process tank below low
level.
Usage exceeding make up rate.
Check feed water supply and (RO)
performance.
Check tank level (visual).
Check take off rate from any
attached equipment, eg. Washers.
COMMS ERROR
Link between keypad and main
PCB faulty
Contact your local distributor or
SUEZ service department.
TEMPERATURE
HIGH
Temperature of purified water
above set limit.
Check temp setting. (Refer to Sec
6.4.2)
Drain off water from system and
leave to refill.
Check feed water temp. (Ref Sec
11.5)
NOTE: If remedial actions do not resolve the problem, turn off the water
supply, isolate the unit from the electrical supply and contact either your
authorised supplier or call SUEZ Service Department. DO NOT ATTEMPT
TO DISMANTLE THE UNIT AND CARRY OUT ANY REPAIRS, UNLESS
TRAINED AND APPROVED TO DO SO, WITHOUT FIRST CONTACTING
SUEZ LIMITED OR YOUR AUTHORISED SUPPLIER.