55
10.2 List of Error Codes
ERROR
CODE
Description
E01
K-type sensor in chamber is out of connection (Emergency light illuminates)
E02
The door is not closed completely (Door light illuminates)
E03
Sterilization temperature over than 137°C (Emergency light illuminates)
E04
The water tank is in LOW WATER condition (LOW-WATER light illuminates)
E05
Water in the chamber is not enough in first 5 min. during add-water step.
E06
Sterilization temperature down under than 2°C over 5 min. in sterilization step
(EMERGENCY light illuminates)
E07
EMERGENCY water/pressure exhaust (EMERGENCY light illuminates)
E08
Jumper open- Error message displayed after vacuum pump operated for 10 min.