A. Position the sterilizer on a stable bench or work surface, ensuring at least 10 cm clearance
between the wall or other pieces of equipment and the sides of the unit for free circulation of
air.
And keep this autoclave in level condition then press the
caster
. To fix the machine.
POWER
SA-300VMA
STURDY
STERI
STERI
ITEM
TEMP
!
Figure 2
B. Pipeline installation:
B-1: Install the water inlet that one end to be fixed at the pipe connector “E” port. And
another end must be fixed on the connector of the water source system.
B-2: Install the exhaust hose: that one end to be fixed on the pipe connector “F” port. And
another end must be fixed on the connector of the water exhaust piping system.
Figure 3
At least 10 cm
Wall
Wall
Wall
At least 10 cm
At least 20 cm
A
:
No Fuse Breaker
C
:
Waste water Filter
D
:
Exhaust for safety valve
E
:
Water inlet
F
:
Waste water outlet
B
:
Power Cord