For REAM mode, rotate the shifter knob
clockwise until it clicks in place.
3. Test the operation of the handpiece.
4. Slide the F/R/Safe control to the safe mode
position until you are ready to use the handpiece.
To Remove Attachment
1. Slide the F/R/Safe control to the safe mode
position.
2. Slide the actuating collar back to release the
attachment.
To Remove Battery Pack
Depress the battery latch and pull the battery pack out
of the handpiece.
To Change Handpiece Mode
WARNINGS:
▪
DO NOT change the handpiece mode while
operating the handpiece.
▪
DO NOT use a reamer in the handpiece and
operate the handpiece in the DRILL mode. Failure
to comply may cause wobble and result in patient
and/or healthcare staff injury.
1. Ensure the handpiece is not running.
2. Change the handpiece mode as follows:
▪
For DRILL mode, rotate the shifter knob
counterclockwise until it clicks in place.
Troubleshooting
WARNING:
DO NOT disassemble or service this equipment.
NOTE:
For service, contact your Stryker sales representative or call Stryker customer service. Outside the US,
contact your nearest Stryker subsidiary.
PROBLEM
CAUSE
ACTION
Handpiece does not run or
turns at a reduced speed.
Battery pack is discharged.
Recharge the battery pack in Stryker battery
charger.
Battery pack is expended.
Replace the battery pack.
F/R/Safe control is in the safe mode
position.
Slide F/R/Safe control to the forward or
reverse mode position.
Drivetrain is malfunctioning.
Return the handpiece for repair.