1. Slide a fully charged battery pack firmly into the
handpiece until the battery latch snaps, indicating
the battery pack is secure (see figure 2).
Figure 2 – To Install Battery Pack
2. Test the operation of the handpiece by sliding the
F/R/Safe Control to FORWARD (F) or REVERSE
(R) and squeezing the trigger.
3. Slide the F/R/Safe control to the safe mode
position until you are ready to use the handpiece.
To Operate Handpiece
WARNINGS:
▪
ALWAYS place the F/R/Safe control in the safe
mode position while the handpiece is idle, before
installing or removing any accessory, or when
passing the handpiece to another person.
▪
Ensure the F/R/Safe control DOES NOT change
position, for example, forward to reverse, while the
handpiece is operating.
▪
A wobbling attachment may cause bone or tissue
damage or inaccurate wiring or pin placement. If
wobbling occurs, see the
Troubleshooting
section.
▪
DO NOT apply excessive pressure, such as
bending or prying, with a cutting accessory
to prevent fracturing the accessory. Applying
excessive pressure, especially during high
operating speeds, may cause the cutting
accessory to bend significantly and result in tissue
damage, loss of tactile control and the ejection of
cutting accessory pieces at a high velocity.
CAUTIONS:
▪
DO NOT stall the handpiece. Failure to comply
may damage the electric motor and/or battery
pack. If the handpiece jams, release the trigger
immediately. Remove any obstructions before
continuing the procedure.
▪
If any power loss is experienced while using a
handpiece, ALWAYS replace the battery pack with
a fully charged battery pack. Failure to comply
may result in a drained or damaged battery pack
with a shortened life.
1. Slide the F/R/Safe control to the appropriate
position to allow the handpiece to operate in the
forward or reverse mode.
2. Squeeze the pressure sensitive trigger for variable
speed operation.
3. Slide the F/R/Safe control to the safe mode
position when you are finished operating the
handpiece.