![Stryker System 6 6203-000-000 Скачать руководство пользователя страница 15](http://html.mh-extra.com/html/stryker/system-6-6203-000-000/system-6-6203-000-000_instructions-for-use-manual_1386046015.webp)
14
www.stryker.com
EN
6203-001-700 Rev-E
Guidance and manufacturer’s declaration - electromagnetic immunity
The System 6 Single Trigger Rotary Handpiece (REF 6203-000-000) is intended for use in the electromagnetic
environment specified below. The customer or the user of the System 6 Single Trigger Rotary Handpiece (REF
6203-000-000) should assure that it is used in such an environment.
Immunity test
IEC 60601 test
level
Compliance
level
Electromagnetic environment - guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80
MHz
3 V/m
80 MHz to 2,5
GHz
N/A
3 V/m
Portable and mobile RF communications equipment should
be used no closer to any part of the System 6 Single
Trigger Rotary Handpiece (REF 6203-000-000), including
cables, than the recommended separation distance
calculated from the equation applicable to the frequency of
the transmitter.
Recommended separation distance:
d
= 1.2√P 80 MHz to 800 MHz
d
= 2.3√P 800 MHz to 2,5 GHz
Where
P
is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and
d
is the recommended separation
distance in meters (m). Field strengths from fixed RF
transmitters, as determined by an electromagnetic site
survey
a
, should be less than the compliance level in each
frequency range
b
.
Interference may occur in the vicinity of equipment marked
with the following symbol:
(Non-ionizing electromagnetic radiation)
NOTE 1:
At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2
: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and
people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM
radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters,
an electromagnetic site survey should be considered. If the measured field strength in the location in which the System 6 Single Trigger Rotary Handpiece
(REF 6203-000-000) is used exceeds the applicable RF compliance level above, the System 6 Single Trigger Rotary Handpiece (REF 6203-000-000) should
be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the
System 6 Single Trigger Rotary Handpiece (REF 6203-000-000).
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.