Do not clean, service, or perform maintenance while the product is in use.
•
Do not steam clean, hose off, or ultrasonically clean the product. Use of these methods of cleaning is not recommended
and may void the product’s warranty.
Recommended disinfectants
•
Quaternaries (2100 ppm active ingredient - ammonium chloride) with no glycol ether
•
Chlorinated bleach solution - 1000 ppm active
•
70% isopropyl alcohol (700,000 ppm)
Recommended disinfection method
1. Follow the disinfectant solution manufacturer’s dilution recommendations.
2. Hand wash all surfaces of the product with a disinfectant solution.
3. Avoid over-saturation and make sure that the product does not stay wet longer than the chemical manufacturer’s
guidelines for proper disinfecting.
4. Dry thoroughly. Do not replace the mattress on the product until the product is dry.
5. Disinfect the hook and loop fasteners after every use. Saturate the hook and loop fasteners with disinfectant, rinse with
water, and allow the disinfectant to evaporate (appropriate disinfectant is determined by the facility).
6. Check functionality before you return the product to service.
•
Raise and lower the product
•
Lock and unlock the brake/steer pedal in both positions
•
Latch and unlatch the siderails
•
Raise and lower the Fowler backrest
•
Make sure all components have proper lubrication
•
Make sure that the jack does not stick due to dust or debris
•
Make sure all labels are intact
N
No
otte
e
•
Direct skin contact with visibly soiled, permeable material may increase the risk of infection.
•
Some cleaning agents are corrosive in nature and may cause damage to the product. If you do not rinse and dry the
product well, a corrosive residue may be left on the surface of the product that could cause premature corrosion of
critical components. Failure to follow these cleaning instructions may void your warranty.
D
Diissiin
nffe
eccttiin
ng
g tth
he
e m
ma
attttrre
essss
W
WA
AR
RN
NIIN
NG
G
•
Always disinfect the mattress following your hospital protocols to avoid the risk of cross-contamination and infection.
•
Do not immerse the mattress in cleaning or disinfectant solutions.
•
Do not allow liquid to pool on the mattress.
•
Do not use V
Viirre
exx®
® T
TB
B to disinfect this product.
•
Do not use accelerated hydrogen peroxides or quaternaries that contain glycol ethers as they may damage the mattress
cover.
EN
32
KK-6300-en Rev 03