Length maximum 90 cm
S
Sp
pe
ecciiffiicca
attiio
on
nss
B
Bo
ottttlle
e ssiizze
e
Diameter 100 to 140 mm / length 465 to 670 mm
3L, 5L
Diameter 140 mm / length 870 mm
UK-F
Diameter 140 mm/ length 900 mm
UK HX
Diameter 140 mm / length 420 to 900 mm
E
Diameter 140 mm / length 420 to 670 mm
C, CD
France 5L, Germany regular 5L O2 bottle, European 5L
For the ST1 non-X-ray model:
•
Diameter maximum 14 cm
•
Length maximum 64 cm
S
Sp
pe
ecciiffiicca
attiio
on
nss
B
Bo
ottttlle
e ssiizze
e
Diameter 100 mm to 140 mm / length 465 mm to 640 mm
3L, 5L
Diameter 100 mm to 140 mm / length 420 mm to 640 mm
C, CD
Germany regular 5L oxygen bottle, European 5L
P
Po
ossiittiio
on
niin
ng
g tth
he
e ttw
wo
o--sstta
ag
ge
e p
pe
errm
ma
an
ne
en
nttllyy a
atttta
acch
he
ed
d IIV
V p
po
olle
e o
op
pttiio
on
n
W
WA
AR
RN
NIIN
NG
G
•
Do not hang IV bags that exceed the safe working load of 18 kg on the IV pole.
•
Do not hang IV bags that exceed the safe working load of 4.5 kg on any hanger on the IV pole.
•
Always make sure that the IV pole is at a low height to pass through door openings when you transport a patient.
•
Do not use the IV pole as a push/pull device. Product damage may occur.
You can purchase the product with the two-stage IV pole option permanently attached at the head end, foot end, or both
ends of the product. The IV pole is equipped with a telescopic pole that extends to provide a second height position. You
can fold and store the IV pole when not in use.
To position the two-stage IV pole (Figure 12):
1. Lift and pivot the pole from the storage position.
2. Push the IV pole down until the IV pole locks in place.
3. To raise the height of the IV pole, pull up on the telescoping portion (A) until the pole locks in place at the highest
position.
4. Rotate the IV hangers (B) to the desired position and hang the IV bags.
5. To lower the IV pole, hold the telescoping portion of the IV pole, turn the latch (C), and lower the telescoping portion.
6. Pull up on the IV pole and pivot the IV pole to the stowed position.
EN
20
KK-6300-en Rev 03
Содержание ST1 Series