Fall Prevention:
110v Outlet
(optional)
A 110v outlet is located under
the footboard at the foot end of
the bed. The 110v outlet has its
own power cord that must be
plugged in.
For Your Safety
The outlet helps reduce the
number of cords strung across
the room, removing a potential
trip hazard.
Only use equipment with the following electrical specs: 110
VAC; 10A; 60Hz. Maximum total load drawn by equipment
used in this receptacle outlet must not exceed 10A. The total
system chassis risk current should not exceed 300uA.
Grounding continuity should be checked periodically. To avoid
risk of electric shock, unplug ALL power cords before opening
the service compartment, junction box or receptacle. Do not
use the optional 110V outlet for life sustraining equipment.
4
Always refer to the package insert, product labeling and/or operations
maintenance manuals before using any Stryker product.