Note:
Pressing the scale button again
within 60 seconds of the first press will
remove the data from the screen.
Nurse Controls: Weighing a Patient
1.
Press and hold the button.
2. “
Release Button
” message flashes on the display.
3. Release
the
Scale
button.
4. After
the
Scale
button has been released, “
Do Not
Touch Bed
” message will flash on the display as shown.
5. When weighing has been completed, the patient’s weight
will be displayed on the status screen as shown. The
patient weight displayed is stored in the system for later
use.
Do not zero bed scales or weigh patient with Percussion, Vibration,
Rotation or Turn Assist active. Patient motion and position resulting
from the dynamic therapy mattress may adversely affect scale
system performance.
The power save mode is activated after one hour on battery power
with no motion release switch activation. Functions including Bed
Exit, Scale and Motion will cease to operate when the unit enters the
power save mode. Injury to the patient could occur if proper patient
monitoring protocol is not observed.
To avoid possible injury and to assure proper operation when using a
powered mattress replacent system such as XPRT:
s
Confirm proper scale system operation following mattress
installation. For best results, secure the mattress power
cord to prevent damage to the cord and interference with
the bed frame and the scale system.
Always refer to the package insert, product labeling and/or operations
maintenance manuals before using any Stryker product.