To Remove the Lid from the Base
1. Push down on the latch release buttons.
2. Slide the latch away from the center of the case to unlock the
latches.
3. Remove the lid from the base.
To Secure the Lid to the Base
1. Place the mats, if applicable, inside the insert tray(s) and stack
the insert tray(s) inside of the base.
2. Put the lid on the base.
3. Slide each latch toward the center of the case until it securely
locks into place.
4. Verify all of the latches are locked and the lid is securely
attached to the base.
Safety Directives
WARNINGS:
▪
Upon initial receipt and before each use, inspect each component
for damage. DO NOT use any equipment if damage is apparent.
▪
Before using this equipment, or any component compatible with
this equipment, read and understand the instructions for use. Pay
particular attention to safety information. Become familiar with the
equipment before use.
▪
Only individuals trained and experienced in the processing of
reusable medical devices should process this equipment.
▪
DO NOT reuse, reprocess, or repackage a device that is intended
for single use only.
-
A single use device may not withstand chemical, chemical
vapor, or high temperature sterilization reprocessing.
-
Design features may make cleaning difficult.
-
Reuse may create a contamination risk and compromise
structural integrity resulting in operational failure or
fragmentation during use.
-
Critical product information may be lost during repackaging.
Failure to comply may lead to infection or cross infection and
result in patient and/or healthcare staff injury.
Maximum Quantity of Devices
WARNING:
DO NOT exceed the specified maximum quantity
of devices in an insert tray. The maximum quantity of devices
is based on Stryker validation of this equipment. Exceeding
the maximum quantity of devices may prevent proper
sterilization of the equipment and/or result in excessive
weight. Always follow the current local recommendations and/
or regulations governing weights and weight limits.
Sterilization Case, Small (REF 6400-276-000)
DEVICE TYPE
INSERT TRAY(S)
BASE
Handpieces
2
1
Attachments
5
0
Accessories
0
1
Total of All Devices
7
2
Sterilization Case, Medium (REF 6400-277-000)
DEVICE TYPE
INSERT TRAY(S)
BASE
Handpieces
5
1
Attachments
6
1
Accessories
0
1
Total of All Devices
11
3
Содержание RemB 6400-276-000
Страница 63: ...6400 277 700 Rev AA EL www stryker com 63 REF 6400 277 000 A B C B D E F REF 6400 278 000 A B C D E...
Страница 68: ...68 www stryker com JA 6400 277 700 Rev AA REF 6400 277 000 A B C B D E F 2 2 4 REF 6400 278 000 A B C D E 3 2 4...
Страница 73: ...6400 277 700 Rev AA ZH www stryker com 73 REF 6400 277 000 A B C B D E F REF 6400 278 000 A B C D E...
Страница 78: ...78 www stryker com KO 6400 277 700 Rev AA REF 6400 277 000 A B C B D E F 2 2 4 REF 6400 278 000 A B C D E 3 2 4...
Страница 82: ......
Страница 83: ......