Fastener assembly weight
46 lb
21 kg
Minimum operators required for loading/unloading an occupied
cot
2
Minimum operators required for loading/unloading an
unoccupied cot
1
Recommended loading height
22 in. to 36 in.
56 cm to 91 cm
Electrical requirements - inductive charging (optional)
12.8 VDC-15.6 VDC, 15A fuse/breaker, 2 conductor 10
AWG cable
Standards
KKK-A-1822F
With inductive charging: IEC 60601-1 Edition 3.0, IEC
60601-1 Edition 3.1, IEC 60601-1-2 Edition 3.0, IEC
60601-1-2 Edition 4.0, IEC 60601-1-12 Edition 1.0,
ANSI/AAMI ES60601-1: 2005/(R)2012, CSA-C22.2 No.
60601-1 (2014)
For standards that require specific options, see
Standards with required options
(page 5).
Stryker reserves the right to change specifications without notice. Patents pending.
The yellow and black color scheme is a proprietary trademark of Stryker Corporation.
E
En
nvviirro
on
nm
me
en
ntta
all cco
on
nd
diittiio
on
nss
O
Op
pe
erra
attiio
on
n
S
Stto
orra
ag
ge
e a
an
nd
d ttrra
an
nssp
po
orrtta
attiio
on
n
Temperature
130 °F
(54 °C)
-30 °F
(-34 °C)
158 °F
(70 °C)
-40 °F
(-40 °C)
Relative humidity
93%
0%
93%
0%
Atmospheric pressure
1060
hPa
620
1060
hPa
620
C
CA
AU
UT
TIIO
ON
N
•
The use of accessories and cables, other than those specified, with the exception of cables that are sold by Stryker as
replacement parts for internal components, may result in increased emissions or decreased immunity of the
P
Pe
errffo
orrm
ma
an
ncce
e--L
LO
OA
AD
D system.
•
Do not use the P
Pe
errffo
orrm
ma
an
ncce
e--L
LO
OA
AD
D system and the P
Po
ow
we
err--P
PR
RO
O cot adjacent to or stacked with other equipment. If
adjacent or stacked use is necessary, observe the P
Pe
errffo
orrm
ma
an
ncce
e--L
LO
OA
AD
D system to confirm normal operation in the
configuration where it will be used.
•
Do not use portable RF communications equipment (including peripherals such as antenna cables and external
antennas) no closer than 30 cm (12 in.) to any part of the P
Pe
errffo
orrm
ma
an
ncce
e--L
LO
OA
AD
D system, including cables specified by
the manufacturer. Otherwise, degradation of the performance of this equipment could result.
•
The emissions characteristics of this equipment make it suitable for use in industrial areas and hospitals (CISPR 11
class A). This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in
accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this
equipment in a residential area (for which CISPR 11 class B is normally required) is likely to cause harmful interference
in which case the user will be required to correct the interference at their expense. In the event of interference, please
relocate or reorient the P
Pe
errffo
orrm
ma
an
ncce
e--L
LO
OA
AD
D system or interfering product.
S
Stta
an
nd
da
arrd
dss w
wiitth
h rre
eq
qu
uiirre
ed
d o
op
pttiio
on
nss
To be compliant with the standards, you must use P
Pe
errffo
orrm
ma
an
ncce
e--L
LO
OA
AD
D with the following compatible cots. See the
Operations Manual for your cot model for more information.
6392-009-002 Rev E.4
5
EN
Содержание Performance-LOAD
Страница 2: ......
Страница 4: ...Fragile handle with care Keep dry 5 Do not stack more than 5 high EN 6392 009 002 Rev E 4 ...
Страница 45: ......
Страница 46: ...Stryker Medical 3800 E Centre Avenue Portage MI 49002 USA 6392 009 002 Rev E 4 WCR AA 5 2018 06 ...