e -- If the tube connector (F) is not at the required depth, use the removed screw (D) to screw into the tube connector
and adjust as required.
F
Fiig
gu
urre
e 8
8 –– A
Ad
djju
ussttiin
ng
g tth
he
e ttu
ub
be
e cco
on
nn
ne
ecctto
orr
4. Reverse steps to reinstall.
N
No
otte
e -- During reinstallation, make sure that you do not overtighten the screws (D) or you will have to readjust the tube
connectors (F).
5. Verify proper operation before returning the product to service.
F
Fo
oo
ott e
en
nd
d cco
ovve
err a
asssse
em
mb
bllyy rre
ep
plla
acce
em
me
en
ntt
T
To
oo
ollss rre
eq
qu
uiirre
ed
d::
•
T27 Torx driver
•
15/16'' hex wrench
•
3/8'' Allen wrench
P
Prro
occe
ed
du
urre
e::
1. Using a T27 Torx driver, remove and save the four socket head cap screws (A) that secure the foot end cover assembly
(B) to the transfer (Figure 9).
2. Using a T27 Torx driver, remove and save the two button head cap screws (C) that secure the foot end cover assembly
to the foot end nose assembly (D) (Figure 9).
3. Remove the foot end cover assembly by sliding it toward the head end of the vehicle patient compartment.
4. Reverse steps 1-3 to reinstall the foot end cover assembly.
6392-009-002 Rev E.4
19