Immunity
Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic
Environment – Guidance
Electrostatic Discharge (ESD)
IEC/EN 61000-4-2
± 8kV Contact
± 15kV Air
± 8kV Contact
± 15kV Air
There are no special requirements with
respect to electrostatic discharge.
Electrical fast transients/bursts
IEC/EN 61000-4-4
Not Applicable
Not Applicable
Surges, line to line
IEC/EN 61000-4-5
Not Applicable
Not Applicable
Surges, line to ground
IEC/EN 61000-4-5
Not Applicable
Not Applicable
Voltage dips, interruptions and
variations on power supply
input lines
IEC/EN 61000-4-11
Not Applicable
Not Applicable
Power Frequency
(50/60Hz) Magnetic Field
IEC/EN 61000-4-8
30A/m
30A/m
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.
There are no special requirements for
non-commercial/non-hospital environments.
Radiated RF
IEC/EN 61000-4-3
10 V/m
80MHz – 2.7GHz
10V/m
80MHz – 2.7GHz
Portable and mobile RF communications
equipment should be used no closer to
any part of the HeartSine samaritan PAD,
including cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of the
transmitter, or 30cm, whichever is greater.
a
Interference may occur in the
vicinity of equipment marked
with this symbol.
Conducted RF
IEC/EN 61000-4-6
Not Applicable
Not Applicable