1-54
28EX-009-002 REV B
En
gli
sh
TRANSFORMER REPLACEMENT (100/200/220/240 VAC OPTIONS)
Tools Required:
•
#2 Phillips Screwdriver
•
1/2” Combination Wrench
Procedure:
Note
Unless otherwise indicated, refer to Figure 14 for this procedure .
1 . Raise the bed to the highest position and apply the brakes .
2 . Unplug the power cord from the wall outlet .
3 . Using a #2 Phillips screwdriver, remove the six screws holding the cover (A) to the litter . Keep the cable clips .
4 . Using a #2 Phillips screwdriver, remove the two screws (A, Figure 13,
) holding the cover plate to the
litter .
5 . Properly ground yourself (see “Grounding Diagram”,
) .
6 . Using a 1/2” combination wrench, remove the nut (B), washer (C) and bolt (D) holding the transformer to the litter .
Disconnect the transformer cable and remove it .
Note
Use caution to avoid overtightening the bolt .
7 . Reverse the above steps to install the new transformer .
8 . Test the bed for proper operation before returning the bed to service .
Service Information
Figure 14
C
B
D
A