1-40
28EX-009-002 REV B
En
gli
sh
FOWLER REPLACEMENT
Tools Required:
•
1/2” Socket
•
(2) 1/2” Combination Wrenches
•
#2 Phillips Screwdriver
•
3/16” Allen Wrench
•
7/16” Combination Wrench
•
Cutting Pliers
•
Strap (or equivalent)
•
3/8” Drive Ratchet
Procedure:
Note
Unless otherwise indicated, refer to Figure 6,
, for this procedure .
WARNING
Do not activate the CPR emergency release handles during this procedure or serious injury to the maintenance technician
or damage to equipment could occur .
1 . Raise the bed to the highest position and apply the brakes .
2 . Unplug the power cord from the wall outlet .
3 . Raise the head siderails .
4 . If applicable, remove the litter plastic cover (if equipped) or raise the fowler to the highest position . Secure the
head section to the bed using a strap to prevent the head section from moving during the procedure .
5 . Loosen the lock rings (A, Figure 1,
) of both siderail cables and remove the siderail cables . Using cutting
pliers, cut the cable ties holding the siderail cables to the frame .
6 . Using a 3/16” Allen wrench and a 1/2” socket, remove the four locknuts and four bolts (B, Figure 1,
)
holding each head siderail assembly to the head section and remove the siderails .
7 . Using a #2 Phillips screwdriver, remove the screw holding each cable tie present on both sides of the fowler
structure .
8 . Using a 1/2” combination wrench, remove the two nuts, two shoulder spacers and two bolts (H) holding the fowler
to the stabilizer arms . Lay the stabilizer arms down .
If the bed is equipped with the CPR mechanism, proceed with steps 9 to 16.
If the bed is not equipped with the CPR mechanism, proceed to steps 17 to 21.
9 . Remove the two wires (C, Figure 17,
) from the micro switch . Note their position before removing them .
10 . Using two 1/2” combination wrenches, remove the locknut, shoulder spacer and bolt (L, Figure 17,
)
holding each CPR handle to the head section . To remove the cable ends from the handle slots, loosen the two
locknuts holding each CPR cable to its handle using a 7/16” combination wrench .
Note
Insert the cable ends inside the handle slots before assembling the handle to the new head section . Then adjust the
two locknuts of each cable in order to have a 1/8” play in the CPR handle movement .
Service Information
Содержание FL28EX
Страница 2: ......
Страница 83: ...www stryker com 28EX 009 002 REV B 1 81 Return To Table of Contents English Notes ...
Страница 99: ...www stryker com 28EX 009 002 REV B 1 97 Return To Table of Contents English Notes ...
Страница 162: ......
Страница 257: ...www stryker com 28EX 009 002 REV B 2 95 Retour à la table des matières Français Commentaires ...
Страница 320: ......
Страница 321: ......