28EX-009-002 REV B
1-13
En
gli
sh
Static Discharge Precautions
The electronic circuits in the GoBed
®
II MedSurg Bed, Model FL28EX are completely protected from static electricity
damage only while the bed is assembled . It is extremely important that all service personnel always use adequate
static protection when servicing the electronic systems of the bed . Whenever you are touching wires, you should be
using static protection .
Static Protection Equipment
The necessary equipment for proper static protection is:
•
1 static wrist strap; 3M part number 2214 or equivalent,
•
1 grounding plug; 3M part number 61038 or equivalent,
•
1 test lead with a banana plug on one end and an alligator clip on the other; Smith part number N132B699 or
equivalent .
CAUTION
All electronic service parts will be shipped in static shielding bags . Do not open the bags until you have completed steps
2 and 3 of the following procedure . Do not place unprotected circuit boards on the floor . All circuit boards to be returned
to Stryker Medical should be shipped in the static shielding bags the new boards were shipped in .
Static Protection Procedure
1 . Unplug the power cord from the wall outlet .
2 . Insert the grounding plug into a properly grounded hospital grade wall outlet . Plug the banana plug of the test lead
into the outlet on the grounding plug . Connect the alligator clip on the other end of the test lead to a ground point
on the bed .
3 . Place the static control wrist strap on your wrist . Connect the alligator clip at the other end of the wrist strap cord
to a ground point on the bed .
BED
GROUNDING DIAGRAM
Содержание FL28EX
Страница 2: ......
Страница 83: ...www stryker com 28EX 009 002 REV B 1 81 Return To Table of Contents English Notes ...
Страница 99: ...www stryker com 28EX 009 002 REV B 1 97 Return To Table of Contents English Notes ...
Страница 162: ......
Страница 257: ...www stryker com 28EX 009 002 REV B 2 95 Retour à la table des matières Français Commentaires ...
Страница 320: ......
Страница 321: ......