Maintenance Manual
42
•
Scale Calibration
If the scale calibration procedure cannot be completed once it has been started, turn the main
switch off and turn it back on to restore the previous calibration data.
N
N
O
O
T
T
E
E
The scale calibration is done in three steps. Steps 2 and 3 come in automatically as soon as a
step is over. Once the calibration process is over, the system returns to the main maintenance
menu.
In the maintenance context, the ZERO key enables the return to the previous level in the
hierarchy of the maintenance menu.
Note that the system will automatically shut off after one minute of idle time when you are at the
main maintenance menu level. Should this happen, the whole process to access the main
maintenance menu will have to be repeated.
Required Tools:
Angle Meter
Procedure:
1. Power off the bed using the main switch at the head end of the bed.
2. Power on the bed and reach the foot end of the bed to simultaneously press both Up/Down
keys of the scale interface menu. This operation allows access to the main maintenance
menu. Display will read:
X
DEBOG MENU
CALIB. MENU
T
N
N
O
O
T
T
E
E
You have exactly five seconds to reach the foot end of the bed and press the keys after having
powered the bed. The system will shut off after this delay, simply repeat the process should it
happen.
3. Choose CALIB. MENU using the Down key and press ENTER. Display will then read:
PUT BED AT 0°
PRESS ENTER
4. Put the angle meter on the seat section of the mattress support and position the bed to 0°.
Then press ENTER. Display will read:
DO NOT TOUCH BED
5. As indicated, do not touch the bed until the end of this step. This is the longest of the three
steps. Once over, the system will bring in the second step. Display will then read:
PUT BED AT +12°
PRESS ENTER
6. Incline the bed in the Trendelenburg position (head down, foot up) until the meter reads +12°.
Ensure that the same material that was on the bed in step 1 stays in the bed. Then press
ENTER. Display will then read:
DO NOT TOUCH BED
7. As indicated, do not touch the bed until the end of this step. Once over, the system will
automatically bring in the third and last step. Display will then read:
PUT BED AT -12°
PRESS ENTER
8. Incline the bed in the Reverse Trendelenburg position (head up, foot down) until the meter
reads -12°. Ensure that the same material that was on the bed in step 2 stays in the bed.
Then press ENTER. Display will then read:
DO NOT TOUCH BED
W
W
A
A
R
R
N
N
I
I
N
N
G
G
Содержание FL25E
Страница 2: ......
Страница 52: ...Appendix A Circuit Diagrams 52 Appendix A Circuit Diagrams A Can Control Board Motor Control Page 1 of 2 ...
Страница 54: ...Appendix A Circuit Diagrams 54 B Power Supply ...
Страница 55: ...Appendix A Circuit Diagrams 55 C Communication Interface Optional ...
Страница 56: ...Appendix A Circuit Diagrams 56 D Scale Optional ...