1-34
1900-009-002 REV B
En
gli
sh
BRAKE ADJUSTMENT PROCEDURE (HYDRAULIC BASE)
TOOLS REQUIRED:
•
3/4 in. Socket
•
3/8 in. Drive Ratchet
•
3/4 in. Combination Wrench
•
(2) Bungee Cords (or equivalent)
PROCEDURE:
1. Raise the litter and the rails to the highest position.
2. Lift the base hood, separating the Velcro fasteners holding it to the base, and support it using bungee cords
(or equivalent).
3. Using a 3/4 in. combination wrench and a 3/4 in. socket, loosen the jam nut (A).
4. Using a 3/4 in. socket, screw in the locknut (B) and test the brakes. Repeat until a proper brake adjustment is
found.
5. Using a 3/4 in. combination wrench and a 3/4 in. socket, tighten the jam nut (A).
6. Remove the bungee cords and replace the base hood.
Service
A
B
Figure 10