EN-35
Guidance and Manufacturer’s Declaration: Electromagnetic Immunity
The Crossfire 2 System is intended for use in the electromagnetic environment specified below. The customer or the user of
Crossfire 2 System should ensure that it is used in such an environment
Immunity Test
IEC 60601 Test Level
Compliance Level
Electromagnetic
Environment - guidance
Electrostatic Discharge (ESD)
IEC61000-4-2
±6kV contact
±8kV air
±6kV contact ±8kV air
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30%.
Electrical fast transient/burst
IEC61000-4-4
±2kV for power
supply lines
±2kV for power supply lines
Mains power quality
should be that of a typical
commercial or hospital
environment
±1kV for input/
output lines
±1kV for input/output lines
Surge IEC61000-4-5
±1kV differential
mode
±1kV differential mode
Mains power quality
should be that of a typical
commercial or hospital
environment
±2kV common mode
±2kV common mode
Voltage dips, short
interruptions and voltage
variations on power supply
input lines IEC61000-4-11
<5% Ut (>95% dip in
Ut) for 0.5 cycle
<5% Ut (>95% dip in Ut) for 0.5 cycle
Mains power quality
should be that of a
typical commercial or
hospital environment.
If the user of Crossfire
2 requires continued
operation during power
mains interruptions, it
is recommended that
Crossfire 2 be powered
from an uninterruptible
power supply or a battery.
40% Ut (60% dip in
Ut) for 5 cycles
40% Ut (60% dip in Ut) for 5 cycles
70% Ut (30% dip in
Ut) for 25 cycles
70% Ut (30% dip in Ut) for 25 cycles
<5% Ut (>95% dip in
Ut) for 5 sec
<5% Ut (>95% dip in Ut) for 5 sec
Power frequency (50/60Hz)
magnetic field IEC 61000-4-8
3 A/m
3 A/m
Power frequency
magnetic fields should
be at levels characteristic
of a typical location in
a typical commercial or
hospital environment.
NOTE: Ut is the a.c. mains voltage prior to application of the test level.
Guidance and Manufacturer’s Declaration--Electromagnetic Immunity
Crossfire 2 System is intended for use in the electromagnetic environment specified below. The customer or the user of
Crossfire 2 System should ensure that it is used in such an environment.
Immunity Test
IEC 60601 Test Level
Compliance Level
Electromagnetic Environment--Guidance
Portable and mobile RF communications equipment
should be used no closer to any part of the Crossfire 2
system, including its cables, than the recommended
separation distance calculated from the equation
applicable to the frequency of the transmitter.
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
3 V
Recommended Separation Distance:
d = 1.2 √P
Содержание Crossfire 2
Страница 1: ...Crossfire 2 Integrated Resection and Energy System 0475 100 000 ...
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 44: ......
Страница 88: ......
Страница 130: ......
Страница 174: ......
Страница 218: ......
Страница 262: ......
Страница 264: ......
Страница 308: ......
Страница 350: ......
Страница 390: ......
Страница 430: ......
Страница 470: ......
Страница 514: ......
Страница 558: ......
Страница 596: ......
Страница 644: ......
Страница 654: ...CHS 650 2 连接交流电源 3 连接机头和脚踏开关 注 释 如果连接了 过期或已使用过的 附 件 控制台 会显示一条错误消息 4 连接抽吸管 用于所有可抽吸设备 ...
Страница 682: ......
Страница 720: ...JA 716 米国製 注意 アメリカ合衆国連 邦法に基づき 本装置は 医師による指示または注 文に応じてのみ販売され ます 壊れ物注意 使用方法を参照してくだ さい ...
Страница 722: ......
Страница 761: ......