Stryker BERTEC GOBED PLUS FL20E Скачать руководство пользователя страница 6

GO

BED

Operations Manual 

 

6

• 

Siderails, with or without their padded covers or nets, are not intended to serve as restraint 
devices to keep patient from exiting the bed. Siderails are designed to keep a patient from 
inadvertently rolling off the bed. It is the responsibility of the attending medical personnel to 
determine the degree of restraint necessary to ensure a patient will remain safely in bed. Failure 
to utilize the siderails properly could result in serious patient injury. 

• 

To reduce risk of injury, ensure the sleep surface is horizontal and in the lowest position with the 
siderails fully raised and locked when moving the bed with a patient in it. 

• 

To avoid injury to the patient and/or user, do not attempt to move the bed laterally with the steer 
pedal engaged. The fifth steer wheel cannot swivel. 

• 

When a patient's condition requires greater safety measures for his/her security, use the lockout 
switches in the foot board control panel to inhibit the siderail functions or remove any optional 
pendant control and install protective pads on the siderails. 

• 

The instant CPR release is for emergency use only. When activating the CPR release handle, all 
people and equipment must be removed from the area below and around the head, thigh and 
foot sections of the bed or serious personal injury or equipment damage could occur. 

• 

Possible fire hazard exists when this bed is used with oxygen administering equipment other 
than nasal, mask type or half bed-length tent type. It is recommended to unplug the bed power 
cord from the wall when oxygen-administering equipment is used. When using a half bed-length 
tent type, ensure the siderails are outside the oxygen tent and oxygen tent should not extend 
below the mattress support level. 

• 

The Weigh system is intended to assist in the monitoring of the patient's weight variation. Under 
no circumstances should its reading be used as sole reference for medical treatment. 

• 

The Bed Exit system  is intended only to aid in the detection of a patient exiting the bed. It is not 
intended to replace patient monitoring protocol. The Bed Exit system signals when a patient is 
about to exit the bed. Adding or removing objects from the bed after having armed the Bed Exit 
system may cause a reduction in the sensitivity of the Bed Exit system, resulting in erroneous 
readings of the patient's movements in the bed. 

• 

The Bed Exit system is not designed to be used with patients weighing less than 50 lb (23 kg). 

• 

When large fluid spills occur in the area of the circuit board, cables and motors, immediately 
unplug the bed power cord from the wall outlet. Remove the patient from the bed and clean up 
the fluid. Have maintenance completely check the bed. Fluids can have an adverse effect on 
operational capabilities of any electrical product. 

DO NOT 

put the bed back into service until it is 

completely dried and has been thoroughly tested for safe operation. 

• 

Do not steam clean, hose off or ultrasonically clean the bed. Do not immerse any part of the bed. 
The internal electrical parts may be damaged by exposure to water. Hand wash regularly all 
surfaces of the bed with warm water and a mild detergent. Wipe cleaned surfaces dry to avoid 
build up of cleaning substance. Inspect the mattress after each use. Discontinue use if any 
cracks or rips are found in the mattress cover, which may allow fluid to enter the mattress. 
Failure to properly clean the mattress, or dispose of it if defective, may increase the risk of 
exposure to pathogenic substances and may bring about diseases to the patient and user. 

• 

Preventive maintenance should be performed at least once a year to ensure all bed features are 
functioning properly. Ensure that any bed malfunction is promptly reported to service personnel 
for immediate attention. 

• 

Always unplug the bed power cord from the wall outlet when servicing or cleaning the bed. When 
working under the bed with the bed in the high position, always place blocks under the mattress 
support frame and apply the brakes to prevent injury in case the Bed Down switch is accidentally 
pressed. 

• 

To avoid damage to the siderail mechanisms, do not move the bed using the raised siderails. 
Use the push/pull handles integrated to the boards to move the bed. 

• 

Before using the optional manual crank during a power failure, always unplug the power cord. An 
unexpected return of power could rotate the handle and cause injury. 

• 

When servicing use only identical replacement parts provided by Stryker Bertec. 

NOTE

 

Throughout this operations manual, the words “right” and “left” refer to the right and left sides of a 
patient lying face up on the bed.

 

Содержание BERTEC GOBED PLUS FL20E

Страница 1: ...Canada E mail Canada only service bertec strykercorp com 1 800 327 0770 United States Manufactured by Stryker Bertec Medical Inc 72 0439 R1 3 September 2003 F15 43 E A Printed in Canada OPERATIONS MANUAL ELECTRIC ACUTE CARE BED Model FL20E USA 2501 ...

Страница 2: ......

Страница 3: ...rm Operation 13 2 6 Patient Restraint Strap Locations 13 2 7 Emergency CPR Release Optional 14 2 8 Night Light Usage Optional 14 2 9 Nurse Call Usage Optional 14 2 10 Auxiliary Power Outlet Usage Option Available w 120V Bed Model 14 2 11 Positioning Siderails 15 2 12 Head and Foot Board Operation 15 2 13 Head Siderail Function Guide 16 2 14 Foot Board Control Panel Guide 17 2 15 Function Lockout U...

Страница 4: ...Height 14 5 to 28 9 36 8 to 73 4 cm Fowler Angle w CPR Mechanism Fowler Angle w o CPR Mechanism 0 to 60 0 to 65 Knee Gatch Angle w o Auto Contour Knee Gatch Angle w Auto Contour 0 to 32 0 to 24 Trendelenburg Reverse Trendelenburg 14 to 14 Environmental Conditions Transport and storage Ambient Temperature Relative humidity Atmospheric Pressure Operating Ambient Temperature Relative humidity Atmosph...

Страница 5: ...mattress thickness should not exceed 6 25 15 9 cm This bed is equipped with a hospital grade plug for protection against shock hazard It must be plugged directly into a properly grounded receptacle Grounding reliability can be achieved only when a hospital grade receptacle is used Serious injury can result if caution is not used when operating the bed Operate the bed only when all people and equip...

Страница 6: ... armed the Bed Exit system may cause a reduction in the sensitivity of the Bed Exit system resulting in erroneous readings of the patient s movements in the bed The Bed Exit system is not designed to be used with patients weighing less than 50 lb 23 kg When large fluid spills occur in the area of the circuit board cables and motors immediately unplug the bed power cord from the wall outlet Remove ...

Страница 7: ... reserves the right to charge shipping and restocking fees on returned items DAMAGED MERCHANDISE Claims for damaged merchandise must be made with the carrier within fifteen 15 days of receipt of merchandise DO NOT ACCEPT DAMAGED SHIPMENTS UNLESS SUCH DAMAGE IS NOTED ON THE DELIVERY RECEIPT AT THE TIME OF RECEIPT Upon prompt notification Stryker Bertec will file a freight claim with the appropriate...

Страница 8: ...ed _____ Depress the green AXIAL STEER side of the pedal at both sides of the bed fully down to put the bed in the steer mode see page 12 Ensure the 5th steer wheel is properly engaged Toggle the pedal to the neutral position and ensure the 5th wheel disengages _____ Ensure the siderails raise lock in the up position and lower smoothly see page 15 _____ Run through each control of the foot board c...

Страница 9: ... mild detergent Wipe the bed clean and dry thoroughly to avoid build up of cleaning solution N NO OT TE E Do not use cleaning solutions containing a degreaser near the siderail pivots see figure 2 2B on page 11 of the Maintenance Manual to avoid deteriorating the grease used to ensure a smooth movement of the siderails Mattress Care Inspect the mattress after each use Discontinue use if any cracks...

Страница 10: ...rol motor resets itself automatically as soon as the fowler is completely lowered ____ Verify the Fowler Knee Gatch and Hi lo movements to ensure that the motion interrupt switch integrated to each of the four actuators is operating properly ____ Auxiliary outlet option available only with 120V electric system beds working properly see page 14 ____ Optional night light working properly ____ Foot p...

Страница 11: ...ITION CONTOUR POSITION Figure 1 9 POWER CORD CONNECTOR AND ON OFF SWITCH MODULE Figure 1 10 QE71 0299 OPTIONAL EMERGENCY CRANK STORAGE LOCATION OPTIONAL 37 PIN NURSE CALL CABLE CONNECTOR OPTIONAL GEN III PENDANT CONTROL CABLE CONNECTOR FOLEY BAG HOOK OPTIONAL NURSE CALL SPEAKERPHONE SIDERAIL INNER CONTROL PANEL FOOT BOARD CONTROL PANEL BRAKE STEER PEDAL CPR RELEASE HANDLE SIDERAIL RELEASE LEVER SI...

Страница 12: ...the red TOTAL BRAKE side of the brake steer pedal at either side of the bed To disengage the wheel brakes toggle the pedal to the neutral position 2 3 MOVING THE BED The bed is equipped with a fifth steer wheel activated by a pedal see page 11 Bed Illustration located at midpoint of the bed on both sides The fifth steer wheel helps guiding the bed along a straight line and helps the bed pivot arou...

Страница 13: ...imple Knee Gatch raise is required To do so perform the following steps 1 Before raising the Knee Gatch reach under the foot section lift the prop rod and hold it while raising the Knee Gatch through the foot board control panel 2 As the Knee Gatch raises the rod will come to pass the support brackets Release it then and continue to raise the Knee Gatch The foot section will then simply follow the...

Страница 14: ...icate through speakerphones with the nurse station by simply pressing a nurse call red button integrated to both head end siderail inner control panels see page 16 The communication between the patient and the nurse station is established the moment a member of the nursing staff responds to the call signal Included with this option are two Ø1 4 phono jacks located on both sides of the head end cas...

Страница 15: ...utward and swing it upward towards the head end of the bed until it latches in the up position To raise a foot end siderail the same procedure is required as for the head end siderail however the siderail swings to the foot end of the bed To lower a siderail lift it up slightly push in the release lever located under the mattress support see page 11 Bed Illustration and hold siderail as it rotates...

Страница 16: ... Siderail Patient s Right Rail Inside Siderail Patient s Left Rail A Press to raise Fowler C Press to raise Knee Gatch B Press to lower Fowler D Press to lower Knee Gatch E Press to activate the Nurse Call function optional F Nurse call speakerphones this panel is optional equipment and will be present on the inner side of both head siderails if the bed is equipped with the Nurse Call option see p...

Страница 17: ...ain the Trendelenburg position head down foot up E2 Press to obtain the reverse Trendelenburg position head up foot down N NO OT TE E To replace the sleep surface to horizontal position after a Trendelenburg positioning simply use the bed height controls to either raise or lower the sleep surface to its limits Then position the bed to the desired height F Contour positioning control The Contour po...

Страница 18: ...splay will read WEIGHT ANGLE XXX X lb XX X Press the MENU Up key to access the Scale Zero mode as written on the membrane The display will read SCALE ZERO PRESS ENTER Press and hold the ENTER key The display will read HOLD TO ZERO WT keep pressing the ENTER key followed by RELEASE TO ZERO release the ENTER key followed by DO NOT TOUCH BED be sure nobody touches the bed while this message appears T...

Страница 19: ... the patient unchanged weight Changing Manually the Numerical Value of Displayed Weight In certain circumstances it may be necessary to change manually the displayed patient s weight To do so Press the key to activate the scale display Access the Change Patient Weight mode using the MENU Down key The display will read CHNG PTNT WEIGHT PRESS ENTER Press and hold the ENTER key The display will read ...

Страница 20: ...may cause a reduction in the sensitivity of the Bed Exit system resulting in erroneous readings of the patient s movements in the bed The Bed Exit system is not designed to be used with patients weighing less than 50 lb 23 kg N NO OT TE E When the bed is equipped with a scale the scale must be properly zeroed for the Bed Exit to function properly see page 18 Operating Zeroing the Scale Before Putt...

Страница 21: ... kg Zone Settings The first zone left LED is the traditional Bed Exit zone The patient can move in the bed freely but cannot fully exit the bed or the alarm will sound The second zone middle LED is more restrictive It allows the patient to sit up and roll over but any attempt to exit the bed will cause the alarm to sound The third zone right LED is the most restrictive zone Small movements like ra...

Страница 22: ... with hook Product Design Three function pendant control Neeco Tron Inc Upright oxygen bottle holder Maximum load 75 lb 34 kg Defibrillator tray Maximum load 40 lb 18 kg Sheet cradle Overhead trapeze system Maximum load 150 lb 68 kg 1 Ø removable anodized aluminium IV pole 1 Ø fixed anodized aluminium IV pole 1 2 Ø removable anodized aluminium IV pole Emergency crank Padded siderail covers set of ...

Отзывы: