Return To Table of Contents
www.stryker.com
3006-409-006 REV C
33
i
Bed® Awareness System Error Codes
Error
Code
Name
Description
Type
Affected Systems
900
Not Used
Message Errors
- When the timeout for the message error has
expired the display should return to the status
screen unless otherwise noted or unless a
higher priority screen is being handled.
901
ScaleSystem
When turning on the Scale system, using the
Scale Button, if a “Logged” error is present
that affects the scale system (see affected
systems column) the message “Function
Disabled - Call Service” will blink.
System
N/A
902
BedExitSystem
When turning on the Bed Exit system if a
“Logged” error is present that affects the
scale or bed exit systems (see affected
systems column) the message “Function
Disabled - Call Service” will blink.
Bed Exit will alarm if Bed Exit is armed and a
200 level exception occurs. The message will
be displayed “Turn off Bed Exit - Call Service”.
When the system is disarmed, the display will
go back to the status screen. Since the bed
exit alarm has the highest priority, if the Bed
Exit alarm is tripped due to the error condition
this message will not be seen.
Message
N/A
903
LBSSystem
When turning on the
i
Bed® Awareness
system if a “Logged” error is present that
affects the
i
Bed® Awareness system (see
affected systems column) the message
“Function Disabled - Call Service” will blink.
i
Bed® Awareness will alarm if
i
Bed®
Awareness is “On” and a 200 or 300 level
Logged error occurs. The message will be
displayed “Turn off Awareness - Call Service”,
When the system is turned off, the display will
go back to the status screen.
Message
N/A
904
UnknownBedType
Unknown bed type occurs when the
configuration switch bank does not contain
a value that is recognized by the software.
This will cause the product to stop all function
and product a message screen indicating the
error number and the message “Unknown
Bed Type - Call Service”
Message
N/A
ERROR MESSAGES (CONTINUED)
NOTE: The 900 error codes are not logged.
Содержание 3002
Страница 2: ......
Страница 82: ...Return To Table of Contents 82 3006 409 006 REV C www stryker com Bed Assembly Standard Bed ...
Страница 83: ...Return To Table of Contents www stryker com 3006 409 006 REV C 83 Bed Assembly Standard Bed ...
Страница 84: ...Return To Table of Contents 84 3006 409 006 REV C www stryker com Bed Assembly Standard Bed ...
Страница 85: ...Return To Table of Contents www stryker com 3006 409 006 REV C 85 Bed Assembly Standard Bed ...
Страница 91: ...Return To Table of Contents www stryker com 3006 409 006 REV C 91 Base Assembly Standard Bed ...
Страница 92: ...Return To Table of Contents 92 3006 409 006 REV C www stryker com Base Assembly Standard Bed ...
Страница 94: ...Return To Table of Contents 94 3006 409 006 REV C www stryker com Typical Both Ends Base Assembly Standard Bed ...
Страница 95: ...Return To Table of Contents www stryker com 3006 409 006 REV C 95 Base Assembly Standard Bed ...
Страница 100: ...Return To Table of Contents 100 3006 409 006 REV C www stryker com Lift Assembly ...
Страница 112: ...Return To Table of Contents 112 3006 409 006 REV C www stryker com Litter Assembly Standard Components ...
Страница 114: ...Return To Table of Contents 114 3006 409 006 REV C www stryker com Litter Assembly Standard Components ...
Страница 117: ...Return To Table of Contents www stryker com 3006 409 006 REV C 117 Litter Assembly Standard Bed ...
Страница 118: ...Return To Table of Contents 118 3006 409 006 REV C www stryker com Litter Assembly Standard Bed ...
Страница 119: ...Return To Table of Contents www stryker com 3006 409 006 REV C 119 Litter Assembly Standard Bed ...
Страница 121: ...Return To Table of Contents www stryker com 3006 409 006 REV C 121 Litter Assembly Standard Bed ...
Страница 127: ...Return To Table of Contents www stryker com 3006 409 006 REV C 127 Electrical Litter Assembly Standard Bed ...
Страница 128: ...Return To Table of Contents 128 3006 409 006 REV C www stryker com Electrical Litter Assembly Standard Bed ...
Страница 129: ...Return To Table of Contents www stryker com 3006 409 006 REV C 129 Electrical Litter Assembly Standard Bed ...
Страница 130: ...Return To Table of Contents 130 3006 409 006 REV C www stryker com Electrical Litter Assembly Standard Bed ...
Страница 134: ...Return To Table of Contents 134 3006 409 006 REV C www stryker com Mechanical Litter Assembly ...
Страница 135: ...Return To Table of Contents www stryker com 3006 409 006 REV C 135 Mechanical Litter Assembly ...
Страница 164: ...Return To Table of Contents 164 3006 409 006 REV C www stryker com Head End Siderail Assembly ...
Страница 165: ...Return To Table of Contents www stryker com 3006 409 006 REV C 165 Head End Siderail Assembly ...
Страница 166: ...Return To Table of Contents 166 3006 409 006 REV C www stryker com Head End Siderail Assembly ...
Страница 167: ...Return To Table of Contents www stryker com 3006 409 006 REV C 167 Head End Siderail Assembly ...
Страница 168: ...Return To Table of Contents 168 3006 409 006 REV C www stryker com Head End Siderail Assembly ...
Страница 189: ...Return To Table of Contents www stryker com 3006 409 006 REV C 189 Foot End Siderail Assembly ...
Страница 190: ...Return To Table of Contents 190 3006 409 006 REV C www stryker com Foot End Siderail Assembly ...
Страница 191: ...Return To Table of Contents www stryker com 3006 409 006 REV C 191 Foot End Siderail Assembly ...
Страница 192: ...Return To Table of Contents 192 3006 409 006 REV C www stryker com Foot End Siderail Assembly ...
Страница 193: ...Return To Table of Contents www stryker com 3006 409 006 REV C 193 Foot End Siderail Assembly ...
Страница 200: ...Return To Table of Contents 200 3006 409 006 REV C www stryker com Notes ...
Страница 228: ...Return To Table of Contents 228 3006 409 006 REV C www stryker com Bed Extender Assembly 3006 346 020 ...
Страница 230: ...Return To Table of Contents 230 3006 409 006 REV C www stryker com Bed Extender Mattress 3000 318 050 ...
Страница 244: ......