D
E
F
G
G
F
Fiig
gu
urre
e 2
20
0 –– S
So
olle
en
no
oiid
d vva
allvve
e
N
No
otte
e -- When you reinstall the bolts (G), tighten the bolts down in an X pattern.
13. Remove and save the solenoid retaining plate (E) from the pump manifold assembly.
14. Unscrew and remove the solenoid valve (D) to be replaced.
N
No
otte
e -- There are six solenoid valves. Five of the six solenoid valves are identical. The bottom right solenoid valve is for
LAL and has a white cap for identification.
15. Plug the pump cord into an accessible outlet.
16. Run the pump diagnostic test. All tests must pass.
17. Reverse steps to reinstall.
18. Verify proper operation before you return the product to service.
P
PC
CB
BA
A a
asssse
em
mb
bllyy rre
ep
plla
acce
em
me
en
ntt
T
To
oo
ollss rre
eq
qu
uiirre
ed
d::
•
T20 Torx driver
•
#2 Phillips screwdriver
•
ESD system
P
Prro
occe
ed
du
urre
e::
1. Unplug the power cord from the wall outlet.
EN
34
2941-009-002 Rev B.0