0702-034-700 Rev-H
EN
www.stryker.com
5
To Reset the Filter Timer
1. Make sure the power cord is connected
between the rover and facility power.
2. Make sure the power switch is in the ON
position and the HIGH SUCTION DEVICE
WARNING has been acknowledged.
3. From the control panel, push and hold
the push button until the SYSTEM SETUP
screen appears (Figure 4).
Figure 4 – To Access System Setup
4. From the SYSTEM SETUP screen, push
the buttons next to the arrow icons to
highlight the SUCTION FILTER LIFE. Push
the button next to OK to make selection
(Figure 5).
OK
ESC
SYSTEM SETUP
VOLUME DISPLAY UNITS
VOLUME DISPLAY BRIGHT
FLUID AUDIBLE INDICATORS
VACUUM DISPLAY BRIGHT
VACUUM DISPLAY CONTRAST
SMOKE FILTER LIFE
SUCTION FILTER LIFE
ERROR HISTORY
Figure 5 – To Select Suction Filter Option
5. The SUCTION FILTER LIFE screen will
appear. Press the display button (RESETS
TIMER) to change the current value of the
filter life to 500 hours (Figure 6).
ESC
SUCTION FILTER LIFE
x / 500
HOURS REMAINING
RESETS TIMER
Figure 6 – To Reset Timer Value
6. Press the display button (OK) to accept
the change. Press the display button
(ESC) to cancel the change (Figure 7).
ESC
OK
SUCTION FILTER LIFE
500 / 500
HOURS REMAINING
TO ACCEPT
OK
Figure 7 – To Accept Timer Value
0000012816, Rev. H Effective Date: Mar 7, 2017 12:56:42 PM