TegraPol-11/ -15
TegraForce-1
Instruction Manual
19
Remove the cap covering the support hole placed on the top of
the TegraPol cabinet.
Remove the two caps covering the fastening screws on the left-
hand side of the TegraPol cabinet by lifting them off with a knife.
Lead the supporting column of TegraForce-1 down into the hole
at the back of the TegraPol.
Then both the distance between disc and specimen mover plate
and the position of the specimen mover plate over the disc have
to be adjusted.
To adjust the correct distance mount a preparation disc on
TegraPol and a specimen mover disc on the TegraForce-1.
Loosen the 2 fastening screws and adjust the space between the
disc and specimen mover plate to about 1 -1.5 mm.
Re-tighten the 2 fastening screws again.
To set the correct position, see Disc Margin in the following
section.
Setting up TegraForce-1 on
TegraPol
Support Hole
Fastening
Screws
Содержание TegraDoser-1
Страница 1: ...TegraPol 11 15 TegraForce 1 Manual No 15517001 Date of Release 0 201 TegraDoser 1 Instruction Manual ...
Страница 6: ......
Страница 50: ...TegraPol 11 15 TegraForce 1 Handbuch Nr 15517001 Auslieferungsdatum 20 TegraDoser 1 Gebrauchsanweisung ...
Страница 101: ...TegraPol 11 15 TegraForce 1 Mode d emploi no 15517001 Date de parution 20 TegraDoser 1 Mode d emploi ...
Страница 106: ......
Страница 152: ...TegraPol 11 15 TegraForce 1 TegraDoser 1 Spare Parts and Diagrams Manual No 15517001 Date of Release 1 20 ...
Страница 159: ......
Страница 160: ......
Страница 162: ......
Страница 163: ......
Страница 164: ......
Страница 165: ......
Страница 166: ......