6. Trouble-Shooting
Display/Error Cause
Action
Error Message
Mounting unit not connected
The plug for the mounting unit is not
connected properly. (Appears when
START
s
is pressed).
Insert the plug firmly. If message still appears,
call a Struers service technician.
PT100 amplifier uncalibrated
The temperature measurement
mechanism is uncalibrated or there is
an error in calibration.
Appears during start-up mode together
with a long beep.
Call a Struers service technician.
Left unit: Pressure Error
Right unit: Pressure Error
Defect in the pressure system.
Appears, when the pre-set pressure is
not reached within 25 seconds.
Call a Struers service technician.
Left unit: Mounting pressure
too high
Right unit: Mounting pressure
too high
Formation of gas in the mounting
cylinder.
Appears, if the actual force should
exceed 65 kN.
Cool the mounting unit down completely.
Move the lower ram down.
Open the top closure.
Error, hydraulic-pump
overloaded
The motor has been overloaded.
Appears, either when pressing: RAM
UP/DOWN
ON
or during the process.
Let the motor cool down and start the mounting
press again.
If this does not help: Call a Struers service
technician.
The plug for the mounting unit is not
correctly connected.
Appears, when the pre-set temperature
is not reached within 6 minutes.
Re-connect the plug.
Left unit: Heating Error
Right unit: Heating Error
Defect in the heating system.
Call a Struers service technician.
Mains tap supplying cooling water is
either closed or not opened sufficiently.
Filter at the water inlet is blocked. The
water is too hot.
Appears, when the cooling is
inadequate.
Open the tap.
Clean the filter.
Insufficient water within the
Recirculation Cooling Unit. The water is
too hot.
Appears, when the cooling is
inadequate.
Fill to the correct water level. See the section:
Maintenance.
Left unit: Insufficient Cooling
Right unit: Insufficient Cooling
Defect in the cooling system.
Call a Struers service technician.
Top closure not unscrewed
The top closure has not been
unscrewed completely after a mounting.
Appears when RAM UP
O
is pressed
after a mounting.
Turn the top closure until it is completely free of
the threads. If this has already been done, the
problem is that the closure cannot spring free.
Hold the swivel arm down with one finger and
press RAM UP
O
again.
Содержание ProntoPress-10
Страница 1: ...Manual No 15077003 Date of Release 11 01 2008 ProntoPress 20 Instruction Manual ...
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 60: ...ProntoPress 20 Gebrauchsanweisung Handbuch Nr 15077003 Auslieferungsdatum 11 01 2008 ...
Страница 61: ......
Страница 63: ......
Страница 65: ......
Страница 121: ...ProntoPress 20 Mode d emploi Mode d emploi no 15077003 Date de parution 11 01 2008 ...
Страница 122: ......
Страница 124: ......
Страница 126: ......
Страница 181: ...Manual No 15077003 Date of Release 10 05 2005 ProntoPress 20 ProntoPress 10 Spare Parts and Diagrams ...
Страница 187: ......
Страница 188: ......
Страница 189: ......
Страница 190: ......
Страница 191: ......
Страница 193: ......