ProntoPress-20
Instruction Manual
11
The cylinder diameter has to be set, in order to have the right
values of force or pressure. When a cylinder diameter is changed,
the database values of force or pressure in the methods will
automatically change to maintain the conditions in the mounting
cylinder.
Note:
The counter which warns the user to clean the rams after
200 mountings, will be reset when a new cylinder diameter is
selected. It is therefore important to clean away any old cured
resin beneath the lower ram when changing cylinder diameter.
Please see the instructions in the ‘Maintenance’ section.
mO
mN
Press MENU UP/DOWN
mON
to select the Cylinder
parameters.
D
CONFIGURATION
Cylinder Diameter: L:
25 mm
R: 50 mm
Temperature Unit: CELSIUS
Force or Pressure: FORCE Unit: kN
D
e
Press ENTER
e
to activate the Cylinder Diameter
Menu.
D
CYLINDER DIAMETER
25 mm
30 mm 40 mm 50 mm
1 1/4" 1 1/2"
D
mO
mN
Press MENU UP/DOWN
mON
to select the diameter
for the cylinder, which is mounted.
D
e
Press ENTER
e
to accept the diameter.
D
Esc
Press Esc to return to the Select Method Group Menu.
Setting the Temperature Unit, Force or Pressure Mode, Force
Unit, Pressure Unit and Countdown Mode: See the section
Advanced Operations.
Setting the Cylinder Diameter
Содержание ProntoPress-10
Страница 1: ...Manual No 15077003 Date of Release 11 01 2008 ProntoPress 20 Instruction Manual ...
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 60: ...ProntoPress 20 Gebrauchsanweisung Handbuch Nr 15077003 Auslieferungsdatum 11 01 2008 ...
Страница 61: ......
Страница 63: ......
Страница 65: ......
Страница 121: ...ProntoPress 20 Mode d emploi Mode d emploi no 15077003 Date de parution 11 01 2008 ...
Страница 122: ......
Страница 124: ......
Страница 126: ......
Страница 181: ...Manual No 15077003 Date of Release 10 05 2005 ProntoPress 20 ProntoPress 10 Spare Parts and Diagrams ...
Страница 187: ......
Страница 188: ......
Страница 189: ......
Страница 190: ......
Страница 191: ......
Страница 193: ......