background image

©Copyright 2021 Strava Solutions, LLC All rights reserved 

PAGE: 7

Initial Set Up/Function Check

NOTE:

 

USE TAP WATER TO FILL UNIT. IF YOU ARE IN A HARD WATER ENVIRONMENT 

WHERE DISTILLED WATER IS PREFERRED, ADD ONE CUP OF TAP WATER TO EVERY 

GALLON OF DISTILLED. UNIT WILL NOT FUNCTION PROPERLY IF DISTILLED WATER IS 

EXCLUSIVELY USED.

 
 

1

 

Plug the Strava supplied power cord (do not use any other 

 

 

power cord other than supplied by Strava) into the back 

 

 

of the unit and a dedicated wall receptacle. (Protective

 

 

Earthing - a dedicated electrical circuit is required). Fill the 

 

 

reservoir with water just above the water sensor 

 

 

(approximately 2”/5cm).  Flip the switch at the back of the 

 

 

unit to turn the unit “on”. At initial powering up of the bath, 

 

 

the start up wizard will display on the LCD screen.

  

 

Start up wizard will prompt user to set time, day of the 

 

 

 

week, and temperature.

  

 

Note: Sanitation cycle defaults to “on”.

 

2

 

Unit must be primed before operation.  Insert the supplied  

 

 

priming bulb into the priming port.  Squeeze the bulb once  

 

 

and hold it until the bulb is fully inflated. Release the tip of  

 

 

the priming bulb from the port.  You should see the water  

 

 

start to bubble at the port.  You will also notice a slight  

 

 

change in the sound of the motor.  If this doesn’t occur,  

 

 

repeat up to 3 times.  Once the water bubbles , the unit is  

 

 

primed.  This process removes any trapped air in the tubes  

 

 

and allows the pump to engage.

 

3

 

Continue filling the tank to desired water level or to the 

 

 

bottom of the internal hinge line.

UNIT INSTRUCTIONS 

WARNING:  Do not start the unit if the cable or plug is damaged, or if there are 
obvious, visible defects to the unit.

Priming bulb

Содержание 100.500

Страница 1: ...dard Size REF 100 500 Rolyan Standard Size REF 100 600E Strava Euro REF 100 500E Rolyan Euro REF 100 600A Strava Aussie REF 100 500A Rolyan Aussie REF 100 600UK Strava UK REF 100 500UK Rolyan UK This...

Страница 2: ...Unit Instructions 7 Initial Set Up Function Check 7 General Use 8 Storage and Handling 9 Error Messages on Control Panel 9 Glass Replacement 10 UV Ballast Replacement 11 Pump Occlusion 13 Heating Ele...

Страница 3: ...the product before using the splint bath The Electric Shock Symbol is used to indicate a hazard arising from dangerous voltage Any mishandling could result in severe injury or death The Exclamation Sy...

Страница 4: ...natural resources and ensure it is recycled in a manner that protects human health and the environment For information about where the user can drop off the waste equipment for recycling please conta...

Страница 5: ...erature Variability F C 2o Sanitation Mode F 185o Set Time Seconds 30 180 Storage Altitude ft 16 732 ft Operation Max ft 16 732 ft m 5100 m m 5100 m C 85o Over Voltage W 11 Frequency Hz 50 60 Hz Mains...

Страница 6: ...ets wheel 8 Temp control up or down 9 Menu 10 Add a new preset 11 Power down unit 4 8 7 9 4 8 7 10 11 2 1 3 5 6 Top Side of Unit Side of Unit 25 5cm 10 60cm 22 625 52cm 19 75 Rear of Unit Inner Tub 30...

Страница 7: ...f the bath the start up wizard will display on the LCD screen Start up wizard will prompt user to set time day of the week and temperature Note Sanitation cycle defaults to on 2 Unit must be primed be...

Страница 8: ...press start Once the bath is empty proceed with tank cleaning Strava recommends a minimum of a monthly water reservoir cleaning more frequently if conditions warrant it Turn the power switch off and u...

Страница 9: ...Carefully remove black rubber boot from UV housing allowing the lamp to slide out of the housing Disconnect lamp from connectors on both ends Replace bulb Reconnect at both ends and insert new bulb ba...

Страница 10: ...rame and ensuring that the glass edge is on track Figure 8 Complete the hardware install by installing the remaining two sex bolt assemblies into the hinges securing all 4 sets with a hand tight Figur...

Страница 11: ...en wrench Figure 15 Lift the scissor drawer out of chassis providing access to the internal control panel Figure 16 Step 2 Remove the old UV ballast Pull out the existing UV bulb from the UV housing a...

Страница 12: ...crew using a 1 Phillips screwdriver Reinsert the UV bulb into the black harness socket Feed the bulb into the UV stainless steel vessel sliding the black rubber grommet over the back of the vessel Fig...

Страница 13: ...ng on the UV stainless steel vessel closest to the back panel Figure 33 Slide flat head screwdriver between the fitting and the hose Push up slowly to slide the silicon hose off of the fitting Figure...

Страница 14: ...ighten with the wire ties aligning them on top of the foam strips of pump important be certain that the wire tie buckles are acclimated to the side of the pump not top Figure 46 47 With pump in one ha...

Страница 15: ...panel Figure 58 Once back panel is removed fold it away from you Remove the two mains screws from the heater with a 2 Phillips screwdriver discard screws at this time Figure 59 60 With the provided 1...

Страница 16: ...the connector Turn the unit on The software upgrade is complete when the set up wizard appears on the screen Step 3 Reassemble the unit Fold the back panel into place aligning the screw holes on eith...

Страница 17: ...d replacement or repair be required purchaser must notify dealer or manufacturer in writing After obtaining an RA returned goods authorization product must be returned in original packaging We estimat...

Страница 18: ...Strava Solutions LLC 2733 Kanasita Drive 125 Hixson TN 37343 USA TEL 1 423 313 6157 EMAIL info kismetus com technicalservice kismetus com www Stravasolutions com MANUFACTURED BY...

Отзывы: