Reprocessing
Instruction For Use • C-MAC
®
video laryngoscope 8401xxx, Electronic Module 8401/8402 X, C-MAC
®
PM 8401
XD • 96076011EN-US • V3.1
23
Clean the optical surfaces of the image chip (8) and LED illuminator (7) on the C-MAC
®
video
laryngoscope with a cotton tip applicator soaked in 70% isopropyl alcohol.
Electrical contacts must be dried very carefully.
7.4 Overview table
Cleaning, disinfection and sterilization
Ultrasound
Manual
cleaning
and disin-
fection
Machine
cleaning
(65 °C)
Machine
cleaning
(93 °C)
Steam ster-
ilization
Gas steril-
ization
Hydrogen
peroxide
sterilization
Steris Sys-
tem 1
®
C-MAC
®
video
laryngo-
scopes
X
●
●
*
X
X
●
●
●
C-MAC
®
PM 8401
XD
X
●
●
*
X
X
X
X
X
Electronic
Module
8401
X/8402 X
X
●
●
*
X
X
●
●
X
●
Suitable methods
X
Unsuitable procedures
*
Device-specific
7.5 Sterilization
CAUTION
Danger due to steam sterilization
Do not steam sterilize the C-MAC
®
video laryngoscope, the Electronic Module 8401 X/8402 X
and the C-MAC
®
PM 8401 XD.
For sterilization with gas or with Steris System1
®
, country-specific laws and regulations
must be observed.
CAUTION
Damage to the product
For sterilization in the Steris System 1
®
and Steris System 1E, the reprocessing cap 8401 YZ
must be put on first.