Installation and commissioning
Instruction For Use • C-MAC
®
video laryngoscope 8401xxx, Electronic Module 8401/8402 X, C-MAC
®
PM 8401
XD • 96076011EN-US • V3.1
15
3
– Residues and/or moisture present at the interfaces
– All components present and correctly positioned
– Mobility of all mobile components, if necessary once the components have been
assembled in the case of instruments that can be dismantled
Test for proper operation
Check the instrument for the following characteristics:
– Fully functional function keys
– Securely positioned connecting cable
– Functional image transmission
– Sufficient image quality
– Light transmission
5.2 Initial operation
WARNING
Optical radiation
The optical radiation of the instrument can injure the eyes of patients, users, and third parties.
Never look into the light output of a connected light transmission or endoscope.
Wear suitable protective equipment.
WARNING
Increased temperatures
The high-intensity light emitted by the laryngoscope can result in increased temperatures at
the light outlet of the LED.
5.2.1 Operating the C-MAC
®
PM
Push the C-MAC
®
PM (12) into the receptacle of the C-MAC
®
video laryngoscope (1) (Fig. A)
Switching on the Open-to-Intubate (OTI) display
1. Fold the OTI display (13) upwards (Fig. A, B).
2. Flip the OTI display to the left (Fig. C).
3. Adjust the OTI display so that the image is optimally visible (Fig. D).
The image focus is adjusted automatically. Manual adjustment is not necessary.