28
Table 202
Guidance and Manufacturer’s Declaration – Electromagnetic Interference Immunity
The charger model 094129/11301 DG is designed for operation in an electromagnetic environment as defined below.
The user of the charger model 094129/11301 DG should make sure that it is used in such an environment.
Interference immunity test
EN°IEC°60601 test level
Conformity level
Electromagnetic environment – guidelines
Electrostatic Discharge (ESD)
IEC 61000-4-2
± 6 kV Contact discharge
± 8 kV Air discharge
Conforms
± 6 kV Contact discharge
± 8 kV Air discharge
Floors should be made of wood or concrete or
provided with ceramic tiles. If the floor is provided with
synthetic material, the relative humidity must be at
least 30%.
Rapid transient electric
interferences /bursts
according to
IEC 61000-4-4
± 2 kV for mains lines
± 1 kV for input and
output lines
Conforms
± 2 kV for mains lines
± 1 kV for input and
output liness
The quality of the supply voltage
should be that of a typical business
or hospital environment.
Surges
according to
IEC 61000-4-5
± 1 kV Push-pull voltage
± 2 kV Common-mode
voltage
Conforms
± 1 kV Push-pull voltage
± 2 kV Common-mode
voltage
The quality of the supply voltage
should be that of a typical business
or hospital environment.
Voltage dips, short-time
interruptions and fluctuations
of the supply voltage
according to IEC 61000-4-11
<5% U
T
*
(>95% dip in U
T
)
for ½ cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
<5% U
T
(>95% dip in U
T
)
for 5 seconds
Conforms
<5% U
T
*
(>95% dip in U
T
)
for ½ cycle
Conforms
40% U
T
(60% dip in U
T
)
for 5 cycles
Conforms
70% U
T
(30% dip in U
T
)
for 25 cycles
Conforms
<5% U
T
(>95% dip in U
T
)
for 5 seconds
The quality of the supply voltage
should be that of a typical business
or hospital environment.
If the user of the device requests continued functioning
even when interruptions of the power supply occur, it
is recommended to supply the device from a power
supply that is free of interruptions.
Magnetic field at the supply
frequency
(50/60 Hz)
according to IEC 61000-4-8
3 A/m
Conforms
3 A/m
The magnetic fields at the mains frequency should
be the values typical of business and hospital
environments.
* Note: U
T
is the mains alternating voltage before applying the test levels.
Anhang
Hinweise zur elektromagnetischen
Verträglichkeit (EMV) für
Ladegerät Modell 094129/11301 DG
Appendix
Information on the electromagnetic
compatibility (EMC) of
charger model 094129/11301 DG
Anexo
Indicaciones relativas a la compatibilidad
electromagnética (CEM) del cargador,
modelo 094129/11301 DG