f) Push the “SKIP TO NEXT STEP” text to activate the step function. The running program will skip directly to the next
step. Remember that each step exist of one ramp and one soak. So, if you push the step button, you will directly jump to
the start of the next ramp.
**Note:
The elapsed soak time will reset when changing from ramp to soak state. This means that if you “STEP” while a soak is
running, and jump to the start of a new ramp, the elapsed soaktime of the previous step will be shown until the ramped
setpoint has reached the target setpoint of that step and start soaking.
h) Push the “CALIBRATE” text to swap to the calibration screen. On the calibration screen the operator can add or
subtract some degrees of the process values to calibrate the controller with your paper trend chard recorder. Touch the
number to set the value by the pop-up keypad. Touch the up and down keys to directly add or subtract one.
i) Push the “PID” text to swap to the PID setting screen. The description of this screen can be found on the Basic settings
chapter “c”.
j) Push the “TREND CHx” text to swap to the channel trend screen, that shows the setpoint and process value of that
specific channel. Touch the screen anywhere to return to the detail channel page.
Password protection: As the calibration screen is password protected, type “USER” in capital letters to enter the
screen. After 1 minute not touching the screen, the password is timed-out and the operator has to type “USER” again if
he wants to enter again of leave the calibration screen.
5) Basic Settings
g) Push the “Manual Hold” text to stop the programmer ramping or soaking. The temperature control will remain on its
current setpoint. Push “Manual Hold” again to
k) From the detail channel screen, push “Back” to return to the detail channel page.