Instruction manual for the hospital bed Deka
Page
38
7
Troubleshooting
7.1
T
ROUBLESHOOTING GUIDE
The following table is a guide for rectifying faults:
PROBLEM
Possible causes
Solution
Manual adjustment
of backrest is not
possible
Rod that connects release
lever and gas spring is bent
or has become detached
Readjust at the release
lever or secure
Manual adjustment
of thigh rest is not
possible
Rod connecting release
lever and gas spring is bent
or has become detached
Readjust at the release
lever or secure
Manual tilting of
mattress base is not
possible
Bowden cable too long /
has become detached
Bowden cable jammed /
kinked
Adjust length of
Bowden cable
Rectify the cause; fix
securely; replace if
necessary
Manual tilting of
mattress base cannot
be blocked
Bowden cable too short
Bowden cable jammed /
does not reset itself
Adjust length of
Bowden cable
Rectify the cause; fix
securely; replace if
necessary