Instruction manual for the hospital bed Deka _____________________________________________
Page
33
6.1.2
Servicing plan
Measure
Material/tool
Check that all screw connections and/or metal
safety caps are firmly positioned;
tighten/replace if necessary.
Suitable tools.
Do not re-use safety caps, to
ensure a perfect hold
If necessary, clean and slightly oil the
connecting bolts on the headboard and
footboard as well as the corresponding sleeves
on the vertical bed frame tubing.
Resin-free and acid-free
grease.
If necessary, clean and lightly spray the pivot
and friction bearings.
Resin-free and acid-free oil
spray (e.g. lubricant A).
Check the adjustment and routing of the
Bowden cable for the tilting of the backrest.
Must move smoothly and easily. No sharp
bends or kinks
Safe routing of Bowden cable.
Touch up any damage in the coating.
Special-purpose paint
(mechanically stable
anticorrosive protection) in a
suitable colour
Special varnishes recommended
Colour Designation
Stiegelmeyer item no.
silver
Varnish (500g tin)
203803
Accelerator (100g tin)
203805
white
Varnish (500g tin)
203804
Key to recommended lubricants
Designation
Stiegelmeyer item no.
A
Weicon: Allround Lubricant AL-W
190507 (400mL aerosol spray)
193231 (1000 mL aerosol spray)
In order to optimise the fault-free operation of the beds and thus minimise
downtime, we recommend that servicing of the bed should include additional
special preventative measures.
A practical guide on preventative replacement of components that may be
affected can be requested under