![Stevens Hydra Probe II Series Скачать руководство пользователя страница 52](http://html1.mh-extra.com/html/stevens/hydra-probe-ii-series/hydra-probe-ii-series_manual_1375858052.webp)
Debug Commands
Warmup Time
Description: Gets/sets the custom probe warmup time.
Access
Level:
Read/Write
Read Addresses: Broadcast, Exact
Read
Command:
<addr>WT=?<CR><LF>
<addr><warmup><CR><LF>
Read Response:
Write Addresses: Exact
Write
Command:
<addr>WT=<warmup><CR><LF>
Write
Response:
<addr><warmup><CR><LF>
Probe Enable
Description:
Gets/sets
whether
the probe circuitry is enabled.
Access
Level:
Read/Write
Read Addresses: Broadcast, Exact
Read
Command:
<addr>PE=?<CR><LF>
<addr><bool><CR><LF>
Read Response:
Write Addresses: Exact
Write
Command:
<addr>PE=<bool><CR><LF>
Write
Response:
<addr><bool><CR><LF>
Measurement Commands
Take Reading
Description: Instructs a probe or group of probes to take a reading sample.
Access
Level:
Execute
Execute Addresses: Broadcast, Exact, Wildcard
Execute
Command:
<addr>TR<CR><LF>
Execute Response:
(none)
Transmit Reading Set
Description: Instructs a probe to transmit a specific reading set.
Access
Level:
Execute
Execute Addresses: Broadcast, Exact
Execute
Command:
<addr>T<set><CR><LF>
<addr><readings><CR><LF>
Execute Response:
52