7-5
Troubleshooting
Operator Manual
920011-649
7.2 Fault Chart
IMPORTANT: When a situation occurs with the Reliance EPS
Processor, an alarm will sound. Refer to S
, on how to respond to faults.
IMPORTANT: When a fault condition or situation occurs, the cycle
automatically stops. Endoscopes may have to be reprocessed
through a complete Reliance EPS Cycle.
FAULT
POSSIBLE CAUSE AND CORRECTION
FAULT 01
– Not used.
Alarm sounds, and display shows:
FAULT 02
CYCLE DELAYED
alternating with:
SEE FAULT CHART
CYCLE DELAYED
Printout message:
HH:MM:SS XM MM-DD-YY
---------------------------------------
= FAULT 02 =
SUMP HTW FILL TOO LONG
NOTE: HTW = Hot Tap Water.
1. Building or processor hot water supply valve is not fully
open.
•
Open hot water supply valves. Press
START
touch
pad to resume cycle.
2. Building supply-line strainers are clogged. Verify strainer;
clean if necessary.
3. Dual Pre-Filter and Pressure Regulator Assembly (required
accessory) needs service.
a. Verify last ten cycle printouts. An increase of maximum
filling time indicates Dual Pre-Filter and Pressure
Regulator Assembly cartridges are clogged. Change
filter cartridges (refer to
S
ECTION
).
b. Perform a cycle. If a situation reoccurs, change
0.2
μm
bacterial-retentive filter cartridge (refer to
S
ECTION
).
c. If situation reoccurs, see
Step 4
.
4. Air bubble in hot water supply lines.
a. Press
START
touch pad to resume cycle. Processor
will be drained and should evacuate any air present.
b. If situation reoccurs, see
Step 5
.
5. Building hot water supply pressure is too low (refer to
equipment drawing [920-011-656]). Call STERIS.*
6. Other possible causes:
- Defective hot water fill valves.
- Defective water level switch.
- Defective drain valve.
7. Call STERIS.*
FAULT 03
– Not used.
*Service charges may be incurred. Consult your warranty for details.
Содержание P920011-649
Страница 1: ...OPERATOR MANUAL Reliance EPS Endoscope Processing System 2022 03 10 P920011 649 ...
Страница 2: ......
Страница 13: ...xi 920011 649 Operator Manual Table of Contents ...
Страница 17: ...xiv 920011 649 Operator Manual List of Tables Healthcare Capital Equipment ...
Страница 25: ...1 8 920011 649 Operator Manual Safety Precautions Healthcare Capital Equipment ...
Страница 45: ...3 16 920011 649 Operator Manual Component Identification Healthcare Capital Equipment ...
Страница 167: ......
Страница 168: ...STERIS Corporation 5960 Heisley Road Mentor OH 44060 1834 USA 440 354 2600 800 548 4873 ...